nonmyd 8    GG-15010

 

Edition 1.1

 

35 

 

Problem 

Cause 

Remedy 

The  image  is 
too dark. 

 

The pupil diameter is small.   

Turn Exposure compensation knob to +1 or +2 position.   
If the patient’s pupil diameter equal to or smaller than that 
shown by the working dots, press SP mode button and 
photograph in SP mode. 

The image is 
out of focus 
or blurred. 

The 

patient 

has 

developed 

cataract. 

Avoiding the area of white turbidity during alignment may 
allow you to obtain a good image. 

The corneal surface is dry. 

Ask the patient to blink before imaging. 

The 

lower 

area  become 
whitened. 

The  eyelid  or  eyelashes  intrude  the 
imaging area. 

Ask the patient to open their eyes wide so that the eyelid 
or eyelashes do not intrude the area of the pupil or you or 
your assistant help the patient keep the eye wide open. 

Flare  appears  on  the 
peripheral area. 

 

The instrument is positioned too 
close to the patient’s eye. 
The  instrument  is  positioned  too  far 
from the patient’s eye. 

Locate  the  camera  to  the  position  where  working  dots 
become  smallest.  When  flares  still  persist,  pull  or  push 
the camera a little.

 

(Working dots may be faded a little.) 

Small pupil photography

SP mode

This  is  not  a  malfunction. 

 

Photography  in  small  pupil 

mode will cause a

 

picture to have flare in its periphery at 

a picture angle of

 

45°. 

Patient's eye with contact lens on. 

This is not a malfunction. 

 

Guide a patient to take off his 

or  her  contact  lenses  from  the  eyes  beforehand  for 
photographing over again.   

Patient  eye  whose  internal  shape 
tends to produce flare. 

A  picture  photographed  may  tend  to  have  flare 
depending

 

on  whose  eye  was  examined.

 

If  the  fundus 

camera should be slightly moved backward

 

and forward 

from  the  position  where  the  fundus  working

 

dots  are 

properly positioned, flare may disappear. 

Internal  diaphragm  for  photography 
dislocated. 

Please contact Kowa or your Kowa dealer. 

A dark shadow appears in 
the center of fundus. 

Insufficient dilation. 

Photograph in the small pupil mode. (Effective pupil 
diameter for photography is 3.3 mm or more.) 
If  the  pupil  diameter  is  too  small  to  photograph  even  in 
the

 

small  pupil  mode,  expedite  a  patient's  pupil  dilation 

by  making

 

the  room  much  darker  or  making  longer  the

 

dark-adaptation time in darkness. 

Black  spots  appear  of  the 
picture photographed. 

Lowpass-filter of digital camera 
contaminated with dust and dirt. 

Clean  the  lowpass-filter  of  a  digital  camera.

 

(Refer  to 

"How to clean up the low-pass filter for Nikon D5300".) 

White spots appear in the 
center 

of 

picture 

photographed. 

Lens  contaminated  with  dust  and 
dirt. 

Clean the objective lens or the lens underneath the hole 
mirror. 

Focus mirror malfunctioning.. 
(Faulty solenoid ) 

Please contact Kowa or your Kowa dealer.After replacing 
the  focus  mirror  solenoid,  mirror  adjustment  will  be 
needed. 

White  spots  appear  at 
places off the center of the 
picture photographed. 

 

Lens contaminated with dirt. 

Clean  the  objective  lens.(  Refer  to  "  Cleaning  the 
objective lens ".) 

Lens with scars on its surface. 

Replace the objective lens. 

Dark shadow or bright 
reflection appears at the 
top or bottom of the 
image. 

 

The instrument is positioned too high 
to the patient’s eye. 

Perform all alignment steps to obtain a clear fundus 
image with consistent brightness. 
Depending on the patient’s eye fixation condition, 
you  may  obtain  a  good  image  by  locating  working  dots 
away from the line. 

The instrument is positioned too low 
to the patient’s eye. 

Dark  shadow  or  bright 
reflection

 

appears  at  the 

left  or  right  of  the

 

image. 

 

The instrument is positioned too right 
or left to the patient’s eye. 

Perform all alignment steps to obtain a clear fundus 
image with consistent brightness. 
Depending  on  the  patient’s  eye  fixation  condition,

 

you 

may obtain a good image by locating working dots away 
from the line. 

Diopter  correction  knob  is  left  in  a 
middle position when switching. 

Check  that  Diopter  correction  knob  is  in  the  correct 
position by slowly moving the knob to see if it cricks. 

 

Содержание D5300

Страница 1: ...nonmyd 8 GG 15010 Edition 1 1 Edition 1 1 2016 11 Model Name K9L61 D5300 model ...

Страница 2: ...lation 20 6 2 The optical head base relation 22 6 3 Power supply relation 23 7 Lens cleaning 7 1 Preparing for lens cleaning 24 7 2 Lens cleaning procedures 25 7 3 Lens cleaning part and cleaning method 26 7 4 Cleaning of objective lens 27 8 Digital camera setting 28 9 Troubleshooting 9 1 Abnormality concerning operation 31 9 2 Abnormality concerning display 33 9 3 Abnormality concerning photograp...

Страница 3: ...nonmyd 8 GG 15010 Edition 1 1 2 1 Name of each component ...

Страница 4: ...nonmyd 8 GG 15010 Edition 1 1 3 ...

Страница 5: ...nonmyd 8 GG 15010 Edition 1 1 4 ...

Страница 6: ...nonmyd 8 GG 15010 Edition 1 1 5 ...

Страница 7: ...nonmyd 8 GG 15010 Edition 1 1 6 ...

Страница 8: ...nonmyd 8 GG 15010 Edition 1 1 7 2 Monitor Screen Indications ...

Страница 9: ...nonmyd 8 GG 15010 Edition 1 1 8 ...

Страница 10: ...nonmyd 8 GG 15010 Edition 1 1 3 Optical pathways ...

Страница 11: ...nonmyd 8 GG 15010 Edition 1 1 10 4 Wiring Diagram ...

Страница 12: ... Stubby type Procedure 1 Remove the caps of both left and right focus knobs with a flat head screwdriver Precision 2 Loosen screws to remove the focus knobs while holding them with your hands to avoid its turning 3 Loosen the screws of flash lamp cover for its removal with a flat head screwdriver 4 Remove the connectors plugged into a digital camera 5 Keep pushing the lens removal button on a digi...

Страница 13: ...ews with phillips screwdriver No 2 Stubby type Remove the front cover 8 Remove four screws with phillips screwdriver No 2 Remove the top cover No 2 9 Remove two screws with phillips screwdriver Remove the right and left cover from the optical head 9 8 ...

Страница 14: ...her optical head base cover 2 Disconnect the cable connector from main board sub board three places Cut a band which holds the cables by using nipper main board side Caution for restoring Never hold the yellow green ground wire There may be a risk of wire disconnection when some load is applied to the ground wire terminal Cut a band which holds the cables by using nipper Disconnect two cable conne...

Страница 15: ...GG 15010 Edition 1 1 14 3 Raise and extract a panel cover And disconnect a cable connector from joystick Be careful not to damage joystick Disconnect a cable connector sub board side Disconnect a cable connector ...

Страница 16: ...he socket Tools Phillips screwdriver No 2 Nipper Procedure 1 Remove the outer cover beforehand See 5 1 2 Remove the optical head base fixing screw with phillips screwdriver No 2 3 Tighten the lock screw 4 Cut a band side of shield box with nipper Remove four screws on shield box with phillips screwdriver No 2 Remove shield box ...

Страница 17: ...ew of the checker Cut a band which holds camera cable and body Cut a band which holds camera cable Disconnect ground wire terminals Disconnect two cable connectors main board Disconnect terminals Disconnect a cable connector Cut a band Disconnect a clamp Disconnect two cable connectors sub board Cut a band Disconnect ground wire terminals Disconnect a clamp Disconnect a cable connector ...

Страница 18: ...nonmyd 8 GG 15010 Edition 1 1 17 7 Lay the optical head base on cushion 8 Removing the power supply and put it on the right side of the main body ...

Страница 19: ...ehand See 5 1 2 Remove the power supply cover See 5 2 3 Remove ten screws Remove the insertion prevention board Raise and extract a panel cover 4 Remove two screws Remove the stoppers left and right 5 Loosen the the lock screw 6 Raise the optical head base And put it next to the power supply cover Note that rack and pinion is kept to be horizontal when returning optical head base to the original p...

Страница 20: ...H Trans H HC K9L57 5440 HV PCB K9L61 6230 SW power supply MUZP 150 12 J0E Interface PCB K9L61 5850 Main PCB K9L61 5150 SUB PCB K9L61 5160 Discharge PCB K9L61 7055A Internal fixation U K9L57 5240 Left side Right side FD LED PCB K9L57 5050H LCD U K9L61 6180 SW R1 PCB U K9L57 7130A SW C PCB K9L57 7133A Lamp VR Flash SW K9L61 7210H SW L1 PCB U K9L58 7132A ...

Страница 21: ...charge board K9L61 5150 K9L61 5160 K9L61 7055A 4 Int Fix unit 5 WD Clean LED 6 AnteLED K9L57 5240 K9L61 7200H HC K9L57 5150 7 FD LED board 8 Digital camera harness 9 ANTE solenoid K9L57 5050H HC K9L57 5300 K9L61 7220H K9L57 5100H 10 Aperture solenoid 11 DIOP sensor 12 LCD unit K9L61 7010H OJ 551 A5 K9L61 6180 ...

Страница 22: ...ion 1 1 21 13 Optical head fan 14 Objective lens 15 MIRROR solenoid HC K9L57 5120 K9L39 9037D K9L39 9037D 16 Obs LED 17 Ring 1 Mono solenoid 18 B W camera HC K9L57 5250 HC K9L61 7190H LOA 19 FD mirror solenoid HC K9L57 5290 ...

Страница 23: ...e relation 1 XeTRG board 2 Optical head motor 3 Menu switch board K9L MU56 2908 HC K9L57 5240 K9L39 9110B 4 Lamp volume and flash switch 5 Joystick 6 Friction tray HC K9L57 5060 ER80 1330 K9L39 750 7 Optical head motor seonsor 8 L R sensor OJ 551 A5 OH 117 A5 ...

Страница 24: ...5010 Edition 1 1 23 6 3 Power supply relation 1 HV board 2 Interface board 3 Power supply board K9L61 6230 K9L61 5850 mUZP 150 12 J0E 4 Power supply fan 5 Capacitor H U 6 TRANS_H K9L57 7808H K9L61 7130 HC K9L57 5440 ...

Страница 25: ...r Other lens cleaners may leave residue after cleaning Softwood stick A softwood stick approximately 10 inches 25 cm in length with its end shaved flat Do not use hard tools such as a flat head screwdriver or a pair of tweezers Hard tools may damage lenses Cleaning paper Gauze may be used instead Do not use chamois leather or silicon cloth Blower One blower is supplied ...

Страница 26: ...otrudes approximately 2 mm off from the tip of stick Dip the tip of cleaning paper into the cleaning solution Clean a lens with a spiral motion from the center towards the rim of the lens Carefully clean the lens surface so that no soiled area is left out Do not apply excess force when cleaning a lens Excess force may damage the lens If the steps above do not clean the lens sufficiently replace th...

Страница 27: ...irror A Cleaning method blower cleaning paper Cleaning part Mask lens Mirror B B W camera lens Cleaning method blower cleaning paper Cleaning part 45 lens Mount protection glass Digital camera Cleaning method blower Refer to How to clean up the low pass filter for Nikon D5300 cleaning paper Objective lens Anterior segment lens Holed mirror Focus lens Mirror A Mirror B Sunspot glass B W camera lens...

Страница 28: ...ing is finished terminate a menu The lamp for the object lens cleaning is turned off A point after having finished wiping it Throw away the cleaning paper gauze used once By wiping with the mixed solution using a new cleaning paper gauze again it can wipe more finely If it wipes while garbage had been attached or it wipes strongly a crack may go into an object lens Please do not use the chamois le...

Страница 29: ...nonmyd 8 GG 15010 Edition 1 1 28 8 Digital camera setting ...

Страница 30: ...nonmyd 8 GG 15010 Edition 1 1 29 For Non UK ...

Страница 31: ...nonmyd 8 GG 15010 Edition 1 1 30 For UK ...

Страница 32: ... improperly connected Reconnect the connector of the chin rest cable and the joystick cable Faulty cable Replace the chin rest cable and the joystick cable Faulty chin rest drive unit Replace the chin rest drive unit The coarse adjustment of the optical head base cannot be smoothly made Wear of friction plate Replace the friction plate Dirty rack pinion Clean the rack pinion Damage of rack pinion ...

Страница 33: ... camera may not be recognized Please use the supplied USB cable Faulty interface board Replace the interface board The right and left position of the optical head is not displayed in the photographic image Connector improperly connected Reconnect the connector of the mini USB cable Faulty cable Replace the mini USB cable Faulty interface board Replace the interface board Fundus observation light c...

Страница 34: ...ime in darkness Connector improperly connected Reconnect the connector of the rotary switch cable Faulty lamp volume or cable Replace the lamp volume flash switch include cable The screen does not reflect in the LCD monitor Contrast brightness maladjustment Adjust the contrast brightness volume in menu Connector improperly connected Reconnect the connector of the LCD cable or coaxial cable Faulty ...

Страница 35: ...to 7 Digital camera setting The flash lamp does not turn on and no image is captured The setting of the digital camera is wrong Confirm the setting of the digital camera Refer to 7 Digital camera setting Connector of trigger cable improperly connected Reconnect the connector of the trigger cable in the correct direction The sign of printed in the trigger cable must turn to the monitor side of the ...

Страница 36: ... appears in the center of fundus Insufficient dilation Photograph in the small pupil mode Effective pupil diameter for photography is 3 3 mm or more If the pupil diameter is too small to photograph even in the small pupil mode expedite a patient s pupil dilation by making the room much darker or making longer the dark adaptation time in darkness Black spots appear of the picture photographed Lowpa...

Страница 37: ...tive when the camera is placed base down as shown at right The assembled D5300 with fundus camera may be cleaning in same way as above 2 Press the menu button to display the menu screen 3 Select Clean image sensor of setup menu and make a selection with right key of multi selector 4 When you press the right key of multi selector again The message shown at right will be displayed while cleaning is ...

Страница 38: ...o Nikon D5300 5 Turn ON the power supply of Nikon D5300 6 Press the menu button to display the menu screen 7 Select Mirror Lock up of setup menu and make a selection with OK button 8 When you press the OK button again the status becomes the ready status for cleaning mirror up 9 As the message When shutter button is turn camera off is displayed press the shutter button 10 The mirror comes up and sh...

Страница 39: ...ad to the front left inspection person side and raise the navigational panel side of an optical head from the bottom Blow away the dust of a right and left detection board from the crevice between a power supply and a mount with Blois or an air duster Twist the dry cleaning paper around ruler And wipe a right and left detection board with the ruler However use neither water nor cleaning solution F...

Страница 40: ...optical head to the front right inspection person side and raise the navigational panel side of an optical head from the bottom Twist the dry cleaning paper around chopsticks And wipe an L R sensor with the chopsticks However use neither water nor cleaning solution L R sensor ...

Страница 41: ...er 2 Disassemble the shading covers after removing screws n 3 3 Clean the lenses 4 Deviate the mirror from optical pathway and hold the mirror to remain the state 5 Set a digital camera of Nikon and then connect each connector Connect nonmyd 8 with filing divice 6 Place the black plane in front of the objective lens When a flash bulb shined align a footstool so that a ring is reflected like an Ima...

Страница 42: ...nob and have 0 positions 10 Loosen screws n 2 a little Do not remove screws from unit please only loose Adjust it by moving Top bottom turn only some monaural sunspot units Tighten a screw after adjustment well Photograph it by pushing the shutter button If most of the photography images become the black photograph it is OK 11 Put back a part in a reverse procedure ...

Страница 43: ......

Страница 44: ......

Отзывы: