Step 1
Position
Position the Sager®
S
301
between the patient’s legs, resting the Ischial
Perineal Cushion (the saddle) against the Ischial Tuberosity, with the shortest end of
the Articulating Base towards the ground.
When positioning, note that the Pulley Wheel should be on the same side and towards the
injured limb. Apply the Abductor Bridle (thigh strap) around the upper thigh of the fractured
limb. Push the Ischial Perineal Cushion gently down while at the same time pulling the thigh
strap laterally under the patient’s thigh. This will seat the lower end of the cushion comfortably
against the Ischial Tuberosity. Tighten the thigh strap lightly. Lift the Spring Clip to extend the
inner shaft until the Pulley (Traction) Wheel is adjacent to the patient’s heels.
Step 2
Set
Note the absence or presence of distal pulses, check for sensation. Position the Malleolar Harness
(ankle harness) beneath the heels(s) and just above the ankle. Fold down the number of comfort
cushions needed to engage all of the ankle above the medial and lateral Malleoli.
Using the attached hook and loop straps wrap the ankle harness around the ankle to secure
snugly. Pull control tabs on the ankle harness to shorten the Ankle sling, pulling it up against
the sole of the foot. Apply
QUANTIFIABLE, DYNAMIC TRACTION™
. Extend the splint shaft to
achieve the amount of traction desired, while observing the amount registered on the Traction
Scale. It is suggested to use 10% of the patient’s body weight per fractured Femur up to 7kg
(15 pounds). At the hollow of the knees, gently slide the large elastic Leg Cravat through and
upwards to the thigh repeating with the smaller cravats to minimize lower and mid-limb
movement.
Step 3
Secure
Adjust the thigh strap at the upper thigh making sure it is not too tight but snug and secure,
then firmly secure the elastic Leg Cravats.
Apply the Pedal Pinion (figure 8 strap) around the feet to prevent rotation. Note the absence
or presence of distal pulses, check for sensation. Patient is now ready for transport.
Warning:
All Operators should receive full and proper initial/refresher instruction sessions from
a qualified person on detailed use of this equipment and regarding the particular situations
in which it should be used.
(PowerPoint #9, 11)
Training application sheet #4:
For model
S
301
Traction Splint.
20
Sager® Emergency Traction Splints and the Minto Fracture Kit