355
Chapter 8
8 .2
Dialog boxes/balloons and countermeasures
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•
The messages displayed on this device are described in the following order:
1) symbols
2) numbers in ascending order
3) alphabet characters in ascending order
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
8 .2 .1 Messages displayed in dialog boxes and countermeasures
If the following messages appear, take the respective countermeasures .
Displayed message
Problems/countermeasures
"[DR Detector name]" requires refresh operation to be used for
next study.
Please refresh the AeroDR by pressing [Refresh] or rebooting
the AeroDR .
[Refresh]: Refresh the AeroDR.
[Cancel]: Return without refreshing.
Displayed when moving to the list screen while there are unre-
ceived images in the DR Detector .
Press [Refresh] to refresh the Detector.
"[ImageAcquisitionConditionName]":Wired/Wireless connection
is changed during image reception .
Return the panel to previous connection . Needs to retrieve the
image .
Displayed when switching a wired or wireless connection while
receiving an image .
Check the connection status of the network device and commu-
nication cables, and tap the Exam. Tags (in the yellow frame) to
re-acquire the image .
[DR Detector name]: "[ImageAcquisitionConditionName]"
Press OK button to retrieve the image .
This panel cannot take an image until the image is fully sent .
Re-acquire the unreceived image. Press the [OK] button.
* Exposure with the corresponding panel is unavailable until im
-
age re-acquisition is complete .
[DR Detector name]: A communication error occurred during
image reception .
Check the connection status of the communication cable .
Displayed when a communication error occurs while the device
is receiving an image from the DR Detector . Check the cable
status. Also, tap the Exam. Tags (in the yellow frame) to re-
acquire the image .
[DR Detector name]: The AeroDR needs to be refreshed.
Do you wish to continue?
An image remains in the panel displayed by a yellow frame on
the Application Bar. Press [OK] to erase images. To recapture
those images without deleting them, open the study that ex
-
posed those images .
[ExposureArea]cm over-sized radiation field detected. Please
set the radiation area again, or add the Exam.Tag.
Displayed when the number of images detected by the AeroDR
Stitching System is greater than the number of Exam. Tags.
Either add Exam. Tags or perform the exposure field detection
again .
[ImageAcquisitionConditionName}: Wired/Wireless connection
is changed during image reception .
Return the panel to previous connection . Needs to retrieve the
image .
Displayed when image reception fails .
Wait until the Detector has returned to the previous connection
and communications are restored .
A barcode is registered to this capture condition key . Is it OK to
change the registration?
Displayed when changing the exposure conditions of a CR
Cassette with a registered barcode to DR Detector exposure
conditions .
Press [OK] to erase the registration.
Continued on the next page
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......