TROUBLESHOOTING GUIDE
If a problem occurs with the Spectrophotometer, please check the following points before
requesting service. If the problem continues to occur even after the suggested corrective
actions have been taken, contact the nearest Konica Minolta-authorized service facility.
Condition
Checkpoint
Recommended action
Refer
to page
Reflectance measurement
results seem strange.
Was specimen
positioned correctly?
Open measurement port
baseplate and check
measurement point.
Are the white calibration
data correct?
Set the correct white
calibration data.
Was white calibration
performed correctly?
Attach White Calibration
Plate correctly and perform
white calibration correctly.
Was zero calibration
performed correctly?
Attach Zero Calibration
Plate correctly and perform
zero calibration correctly.
Is there any obstacle in the
transmittance chamber?
Make sure that there is nothing
in the transmittance chamber.
Was measurement mode
set to reflectance?
By referring to the operation
manual of the software, perform
the operation correctly.
—
Transmittance
measurement results
seem strange.
Was specimen
positioned correctly?
Position specimen correctly.
Was 100% calibration
performed correctly?
Attach White Calibration Plate
correctly and perform 100%
calibration correctly using
the appropriate calibration
standard (air or distilled water).
Was zero calibration
performed correctly?
Attach White Calibration
Plate correctly, block all light
from reaching the receptor
window, and perform zero
calibration correctly.
Was measurement mode
set to transmittance?
By referring to the operation
manual of the software, perform
the operation correctly.
—
Data input/output between
the Spectrophotometer
and a computer cannot
be performed.
Commands cannot be input
to the Spectrophotometer
from a computer.
Is the USB cable connected
correctly to both the
Spectrophotometer
and the computer?
Connect the USB cable correctly
bitween the Spectrophotometer
and the computer.
I s t h e s o f t wa r e o p e ra t i n g
correctly?
By referring to the operation
manual of the software, perform
the operation correctly.
—
Set POWER switch of
Spectrophotometer to OFF
and then set it back to ON.
Содержание CM-3700A
Страница 1: ...SPECTROPHOTOMETER CM 3700A INSTRUCTION MANUAL Before using this instrument please read this manual ...
Страница 9: ...7 Using the CM 3700A ...
Страница 34: ...32 ...
Страница 35: ...33 Explanations ...
Страница 39: ...37 Dimensions 170 6 170 6 37 27 247 6 75 5 91 5 197 371 271 2 7 496 478 71 8 91 5 mm ...
Страница 42: ... 2011 2017 KONICA MINOLTA INC 9222 A4KX 11 BHEEDK ...