24
Measurement Functions of Probes
1. Measurement Speeds
A probe has the following four modes (measurement speeds), which are available according to the purpose.
A probe outputs every measured value and a PC or data processor connected to it can retrieve all of them through
communication. Generally, however, not all measured values are displayed but values are displayed at an appropriate
interval.
FAST mode
Requires short measurement time, but may compromise measurement repeatability for low-luminance measurements.
SLOW mode
Repeats FAST mode measurements several times. Use this mode when you want to perform measurement with high
repeatability.
AUTO mode
Switches FAST, SLOW, and setting with a further longer integration time automatically depending on the luminance level.
Normally, this is the recommended setting.
LTD.AUTO mode
Switches between FAST and SLOW settings automatically depending on the luminance level.
Notes on selecting the measurement speed
•
The currently selected measurement speed setting is stored in the probe.
The set measurement speed will be used at the next start-up.
Содержание CA-410 Series
Страница 12: ......
Страница 40: ......
Страница 139: ......
Страница 140: ...En 9222 AA1J 12 2017 KONICA MINOLTA INC BIFBDA...