RS-232 serial line connection using three wires
An RS-232 serial line connection using three (3) wires and NMEA datagrams is a
common way to connect the EK80 to external devices.
A
Local connection
B
Connection on remote device
C
Female 9-pin D-Subminiature connector
D
Male 9-pin D-Subminiature connector
Unless otherwise specified, this cable must be provided
by the installation shipyard. Note that this cable does
not support all the signals in the standard RS-232
specification.
Minimum cable requirements
•
Conductors
: 2 x 4 x 0.5 mm
2
•
Screen
: Overall braided
•
Voltage
: 60 V
•
Maximum outer diameter
: Defined by the plugs and/or the cable gland
If you need to install a very long cable, increase the cross section.
Related topics
Cable drawings and specifications, page 136
Cable layout and interconnections, page 85
Simrad EK80 Installation Manual
Содержание Simrad EK80
Страница 1: ...kongsberg com simrad Simrad EK80 INSTALLATION MANUAL ...
Страница 2: ......
Страница 12: ...10 394149 D Simrad EK80 ...
Страница 223: ...394149 D 221 6 Start the installation by selecting Next 7 Select Finish to close the wizard Setting to work ...
Страница 313: ...394149 D 311 Drawing file ...
Страница 314: ...312 394149 D Simrad EK80 Installation Manual ...
Страница 317: ...394149 D 315 Drawing file ...
Страница 318: ...316 394149 D Simrad EK80 Installation Manual ...
Страница 320: ...318 394149 D 201575 Transducer connector assembly and wiring Simrad EK80 Installation Manual ...
Страница 352: ... 2020 Kongsberg Maritime ISBN 978 82 8066 182 1 ...
Страница 353: ......
Страница 354: ...Installation Manual Simrad EK80 ...