Kofloc FML-500 Скачать руководство пользователя страница 6

                                                                         

MJ400252C1   

 

- 5 - 

4.

 

Specifications 

 
Standard specifications 

Measuring fluids 

Water, pure water (bubbles not included)    *For other 
liquids, please contact us. 

Fluid sonic range 

1300 m/s 

  1600 m/s 

Fluid temperature range 

15 

  50

 

Measuring range 

  5 L/min 

Measuring accuracy 

See Table 1. 

Materials of parts in 
contact with liquid 

New PFA 

Pressure resistance 

0.5 MPa

at 20

℃)

 

Process connection 

6-mm tube end 

Cable length 

3 m 

Cable sheath 

PVC 

Enclosure 

Equivalent to IP64 

Ambient temperature 

  50

℃(

No condensation

 

Ambient humidity 

10 

  85%RH max. (No condensation) 

Installation 

Vertical, horizontal, inclined 

 

   

Table 1: Measuring accuracy (water 23

Measuring range  0 

  5 L/min 

Flow range 

 0.20 L/min 

  0 

 5.00 L/min 

Measuring 
accuracy 

F

lo

w

 

ve

lo

cit

Below 1 m/s (below 0.8 L/min

 

±30 mL/min 

  2 m/s

0.8 

  1.5 L/min

 

±2.5%F.S 

  3 m/s

1.5 

  2.2 L/min

 

±2.0%F.S 

  6.6 m/s

2.3 

  5 L/min

 

±1.5%F.S 

 

   

Electrical specifications 

Measuring method 

Ultrasonic wave propagation time difference calculation 

Output   

Flow output 

 20 mA

Load resistance 500 

Ω

 max.

 

Flow alarm 
output 

Open collector output 2 points 
Load rating: 30 VDC, 20 mA, operation: comparison 

Indicator 

4-digit, 7-segment LEDs 

Indication 

Instantaneous flow rate, integrated flow rate 

Alarm output monitor 

LED 2-point indication (Instantaneous flow rate alarm, 
integrated flow rate alarm) 

Parameter setting 

Four key switches on the panel 

Power supply 

24 VDC±5% 

Power consumption 

150 mA max. 

 

 

 

Содержание FML-500

Страница 1: ...MJ400252C1 Straight Type Ultrasonic Flow Meter FML 500 Instruction Manual...

Страница 2: ...improvement Prior to shipment every care has been taken in preparing this manual not to mention the product itself but if you notice any imperfections errors or omission please contact KOFLOC Prior to...

Страница 3: ...3 3 Overview 4 4 Specifications 5 5 Installation 5 1 Installation place 7 5 2 Installation method 7 5 3 Precautions for piping 7 6 How to Use 6 1 Warming up 9 6 2 Operating procedure 9 6 3 Description...

Страница 4: ...LOC to the installation site in order to prevent damage due to accidents during transportation 2 3 Precautions for storage of the product If the product is not put in use for a long time after it was...

Страница 5: ...liquid made of New PFA Straight type passage to cause no unnecessary pressure loss and facilitate maintenance Small size and light weight Compact Meter equipped with an indicator can be made ready fo...

Страница 6: ...easuring accuracy water 23 Measuring range 0 5 L min Flow range 0 0 20 L min 0 5 00 L min Measuring accuracy Flow velocity Below 1 m s below 0 8 L min 30 mL min 1 2 m s 0 8 1 5 L min 2 5 F S 2 3 m s 1...

Страница 7: ...External dimensions Signal table Pin No Signal Name Core Color 1 Power supply 24V Red 2 Power supply current output COM Black 3 Current output White 4 Alarm output 1 Green 5 Alarm output 2 Yellow 6 Al...

Страница 8: ...the equipment in a condition deviating from the conditions described in the specifications is a cause of failure 5 2 Installation method Install the meter using the threaded holes M3 on the bottom of...

Страница 9: ...product exercise caution not to damage the inside If it is damaged the accuracy will deteriorate Check to see if the connections of piping are not leaking If such a check is omitted and unsafe liquid...

Страница 10: ...being indicated IF lights and when an integrated flow rate is being indicated TF lights Shift key Measurement Switches over the indication of flow rates between the instantaneous flow rate and the int...

Страница 11: ...shing WIN1 Re measure mode selected Flashing W Re measurement being executed Flashing Character move AVR0 Auto AVR selected Flashing AVR1 Auto AVR selected Factory set value Flashing R Auto AVR being...

Страница 12: ...ed flow rate are the units set in each setting mode Integrated flow rate resetting In the state of integrated flow rate indication press and hold the up key and down key for approx 3 seconds Then the...

Страница 13: ...ate or integrated flow rate Default values Mode No Default Value Set Item ABCD A B C D P 01 1 0 Damping constant setting P 02 3 Unit of flow rate setting 0 5 0 0 Full scale flow rate setting P 03 Fact...

Страница 14: ...ll scale flow rate setting A B C D 0 5 0 0 Input range mL s 3 14 99 99 mL min 188 9999 L min 0 18 11 30 L h 11 678 After a full scale flow rate has been set when YES is selected with the up key the va...

Страница 15: ...t range 0 00 9 99 g cm3 Mode No Liquid kinematic viscosity coefficient setting P 08 A B C D 0 1 0 0 Set range 0 00 99 99 cSt Mode No Upper limit flow rate alarm setting P 09 A B C D 0 5 0 0 Set range...

Страница 16: ...imit alarm 4 Integrated flow rate alarm 5 No received wave alarm Mode No Alarm output 2 type assignment setting P 14 A B C D 2 0 No alarm output 1 Instantaneous flow rate upper limit alarm 2 Instantan...

Страница 17: ...upward in order to prevent trapping of air bubbles b Measuring liquid contains bubbles or foreign matter 1 Eliminate bubbles and foreign matter as much as possible by for example installing a deaerato...

Страница 18: ...hine occurs due to the following reasons even within the warrant period it becomes the outside for a warranty a Malfunctions due to erroneous applications repairs or remodeling Including the case in w...

Страница 19: ...MJ400252C1 18 URL http www kofloc co jp...

Отзывы: