K8853X Air Release Value Instrument
Operation and Instruction Manual
K8853X-Manual
-
15-
8 Troubleshooting
WARNING:
Troubleshooting procedures
involve working with high voltages and/or
temperatures which may result in personal injury
or death, and should only be performed by trained
personnel. Please do not hesitate to contact
Koehler for assistance.
9.1 Unit does not power up
1. Establish that the socket outlet is providing
proper and adequate voltage.
2. Check if Overpower Protection circuitry
located directly behind the temperature
controller inside the front tray has been
activated.
3. Check if line switch is in the
ON
position.
4. If problem persists, please call the Koehler
technical service department for assistance.
9.2 Unit is on and keeps resetting into
start up routine
•
For 220V units, make sure that the socket
outlet is greater than 215V.
•
Check if there is a steady and reliable power
source.
•
Make sure the connector plug on the rear
panel is firmly plugged in.
9
Service
Under normal operating conditions and with
routine maintenance, the K8853X Air Release
Value Instrument should not require service. Any
service problem can be quickly resolved by
contacting Koehler’s technical service department
either by letter, phone, fax, or email. In order to
assure the fastest possible service, please provide
us with the following information.
Model Number:
Serial Number:
Date of Shipment:
10 Storage
This laboratory test instrument is equipped with
electrical components. Storage facilities should be
consistent with an indoor laboratory environment.
This testing equipment should not be subjected to
extremes of temperature and/or moisture.
This equipment was shipped from the factory in
a corrugated cardboard container. If long term
storage is anticipated, re-packing the instrument in
a water-resistant container is recommended to
ensure equipment safety and longevity.
11 Disposal
11.1 General Recycling Information
•
No refrigerants or pressurized materials.
•
No charged capacitors or components that
could electrically discharge.
•
No components such as springs or spring
powered gears that could store mechanical
energy.
•
No chemical hazards from any components.
•
No radiation is emitted from any components.
11.2 Disposal Information
The K8853X Air Release Apparatus is Rohs
compliant and is subject to the WEEE directive.
Once the instrument has reached the end of its
useful life, the instrument and or its components
must be recycled or disposed of in accordance to
Country, State, or local laws that may impose
regulatory
requirements
regarding
disposal.
Dispose
instrument
and/or
components
in
accordance to all applicable regulations.
The K8853X is composed of the following major
components and materials:
Major Component
Material
Instrument Housing
Powder Coated, Cold-
Rolled Carbon Steel 18
Supports and Racks
Stainless Steel, #316
Heater Coils
Copper
NOTE:
There are No Refrigerants or Hazardous
Materials contained in any component.