K463XX Cloud and Pour Point Bath 

Operation and Instruction Manual 

 

K463XX-Manual 

- 12 - 

 

Setting  the  Temperature

.  Set  the  desired 

operating  temperature  by  adjusting  the  set 
point  with  the  up  and  down  keys.  The  set 
point will be displayed in the lower green Set 
Point 

LED 

display 

and 

the 

actual 

temperature  will  be  displayed  in  the  upper 
red  Process  LED  display.  Please  allow  the 
instrument 

to 

fully 

equilibrate 

before 

proceeding with any testing. 

 

Temperature  Calibration.

  This  routine  allows 

the  digital  temperature  controller  to  be 
calibrated  to  a  certified  thermometer.  In  the 
calibration mode, the display will always give 
2 decimal places. 

 

a.  Use  a  certified  calibrated  measuring 

device  to  acquire  the  temperature. 
Calculate  the  difference  between  the 
measuring  device  and  the  Process 
value displayed on the controller. 

b.  Press  the  program  key  two  times  until 

PCt

 is displayed in the lower green LED 

display.  Press  the  DOWN  key.  CAL  will 
display  on  the  lower  green  display.  If 
there  is  a  value  observed  in  the  upper 
red LED display, add it to the calculated 
difference obtained in the previous step. 
This is the offset value. 

 

c.  Press  the  Program  Key.  The  lower 

green  display  will  flash.  Use  the  up  or 
down  keys  to  adjust  to  the  new 
calibration offset value  on the upper red 
display  calculated  in  the  previous  step. 
When  the  value  has  been  entered,  the 
controller  will  automatically  store  the 
value.  The  lower  green  display  will  stop 
flashing.  If  further  adjustments  are 
necessary,  press  the  Program  Key 
again.  Resume  regular  operations  by 
pressing  the  Exit  /  Function  key  two 
times.  Verify  if  the  new  calibration  is 
correct  by  observing  the  upper  red 
display  and  comparing  the  value  with 
the calibrated reference device.  

 
 

Auto  Tune.

  This  routine  allows  the  digital 

temperature  control  to  learn  the  heating 
parameters  needed  for  any  particular  set 
point temperature. This  operation  should  be 
done  when  installing  a  new  unit,  after 
replacing or changing the bath medium type, 
or  utilizing  a  different  temperature  set  point 

20%  different  from  the  previously  used  set 
point temperature. 

 

a.  Set  the  operating  temperature  to  the 

desired setting. 

 

Press the up and down arrow buttons 
simultaneously for about 5 seconds. When Auto 
Tune is active, the lower green LED display will 
blink 

TUNE

. Auto Tune will automatically toggle 

off when the set point temperature is reached. 
Auto tune can be terminated by pressing the up 
& down buttons simultaneously again.

 

Operation 

Be  sure  to  read  the  safety  and  hazard 
warnings, the installation procedure and any 
of  the  standard  test  methods  mentioned  in 
the  Introduction  of  this  manual  before 
operating this unit. 

5.1  Bath 

Fill  each  bath  chamber  with  the  appropriate 

bath  medium.  This  can  be  done  by  removing 
one  of  the  copper  jackets  in  each  chamber. 
Allowance  must  be  given  for  the  liquid 
displacement that  would occur when the copper 
jacket is returned to its position. When all jackets 
are in position, the liquid medium level should be 
no more than 25 mm from the top. The level can 
be checked my means of a dipstick through the 
thermometer hole. 
 
The manufacturer is unable to make any specific 
recommendation  on  the  choice  of  a  bath 
medium.  Safety,  toxicity,  flammability  and  other 
factors  must  be  taken  into  account  while 
selecting  a  bath  medium. However,  during  tests 
at  the  factory,  200  proof  (water  free)  denatured 
ethanol was used with satisfactory results. Other 
users  have  reported  methanol  to  also  be  an 
adequate bath medium as well. 
 
Glycol  water  mixtures  are  not  a  very  suitable 
bath  medium  since  partial  freezing  around  the 
refrigeration  coils  and  high  fluid  viscosity  will 
reduce  cooling  efficiency  with  the  resultant 
increase  in  bath  temperature.  Furthermore,  in 
extreme  cases,  damage  to  the  bath  may  also 
occur which will void the warranty. 
 
 
 

Содержание K463XX

Страница 1: ...nue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis I...

Страница 2: ......

Страница 3: ...K463XX Cloud and Pour Point Bath Floor Model Operation and Instruction Manual K463XX Manual 1...

Страница 4: ...K463XX Cloud and Pour Point Bath Floor Model Operation and Instruction Manual K463XX Manual 2...

Страница 5: ...ance with the applicable standards set forth in this certification Specifications ASTM D97 ASTM D2500 ASTM D5853 ASTM D6074 ASTM D6158 IP 15 IP 219 ISO 3015 ISO 3016 DIN 51597 FTM 791 201 NF T 60 105...

Страница 6: ...K463XX Cloud and Pour Point Bath Floor Model Operation and Instruction Manual K463XX Manual 4...

Страница 7: ...information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Rep...

Страница 8: ...ately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is...

Страница 9: ...IFICATIONS 9 2 SAFETY INFORMATION AND WARNINGS 9 3 GETTING STARTED 10 3 1 PACKING LIST 10 3 2 UNPACKING 10 4 DESCRIPTIONS 11 4 1 INSTRUMENT CONTROLS 11 4 2 ADDITIONAL ACCESSORIES 11 4 3 TEMPERATURE CO...

Страница 10: ...our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the...

Страница 11: ...pment Modifications and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety perform...

Страница 12: ...floor Individuals must lift in accordance to proper technique See Figure below Unwrap the unit and wipe the shell with a soft cloth to remove smudges caused by wrapping material Very high ambient tem...

Страница 13: ...probe 2 Set Point Temperature Display The lower green LED display shows the set point temperature of the controller 3 Programming Key Permits scrolling through controller menu parameters One Level Fo...

Страница 14: ...he heating parameters needed for any particular set point temperature This operation should be done when installing a new unit after replacing or changing the bath medium type or utilizing a different...

Страница 15: ...ved because there is a possibility of electrical shock 4 Proceed to test in accordance with published test methods see Section 1 3 6 Maintenance WARNING Disconnect power to the unit before servicing t...

Страница 16: ...K463XX Cloud and Pour Point Bath Operation and Instruction Manual K463XX Manual 14 7 Wiring Diagram...

Страница 17: ...ument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive r...

Страница 18: ...K463XX Cloud and Pour Point Bath Operation and Instruction Manual K463XX Manual 16 Notes...

Страница 19: ...K463XX Cloud and Pour Point Bath Operation and Instruction Manual K463XX Manual 17 Notes...

Отзывы: