background image

 

 

 

 
 

CERTIFICATE OF CONFORMANCE 

 

Low Temperature Brookfield Viscosity Bath, BVS3000 

K34710, K34711, K34712 

 

 

This  certificate  verifies  that  part  numbers  K34710,  K34711,  K34712  Low 
Temperature Brookfield Viscosity Bath, were manufactured in conformance 
with the applicable standards set forth in this certification. 
 
 

Specifications:  

 

 

ASTM D2983 
IP 267 Method A 
ISO 9262 
CEC-L-18A 

 
This unit is tested before it leaves the factory, to ensure total functionality 
and compliance to the above specifications and ASTM standards. Test and 
inspection records are on file for verification

 
 

 

                

Jesse Kelly  

 

 

 

 

 

      

Application Engineer 
Koehler Instrument Company 

 

Содержание K34710

Страница 1: ...Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing Brookfield Viscosity Liquid Bath K34710 K34711 K34712 ...

Страница 2: ......

Страница 3: ...e manufactured in conformance with the applicable standards set forth in this certification Specifications ASTM D2983 IP 267 Method A ISO 9262 CEC L 18A This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler Instrume...

Страница 4: ......

Страница 5: ...ent Specifications 3 2 Safety Information and Warnings 3 3 Unpacking Installation 4 3 1 Packing List 4 3 2 Unpacking 4 3 3 Instrument Installation 4 4 Instrument Descriptions 5 4 1 Temperature Controller Operation 6 5 Operation 7 5 1 Viscosity Bath 7 5 2 Testing Procedure 7 6 Safety Features 7 6 1 Over Power Protection 7 7 Maintenance 7 7 1 Routine Maintenance 7 7 2 Replacement Parts 7 8 Service 8...

Страница 6: ... 1 2 1 1 Koehler s Commitment to Our Customers Providing quality testing instrumentation and technical support services for research and testing laboratories has been our specialty for more than 50 years At Koehler the primary focus of our business is providing you with the full support of your laboratory testing needs Our products are backed by our staff of technically knowledgeable trained speci...

Страница 7: ...ecommended voids the manufacturer warranty product safety performance specifications and or certifications whether specified or implied and may result in personal injury and or property loss Replacement parts must be O E M exact replacement equipment Unit Design This equipment is specifically designed for use in accordance with the applicable standard test methods listed in section 1 2 of the oper...

Страница 8: ...nspect the unit and all accessories for damage If you find any damage keep all packing materials and immediately report the damage to the carrier We will assist you with your claim if requested When submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment Do not return goods to Koehler without written authorization 3 3 Instrument Installation The...

Страница 9: ...fety please disconnect the line cords whenever performing any maintenance operations and or cleaning of the unit WARNING Do NOT turn the power on unless the bath is filled with the proper medium Low liquid level exposes the heater which may ignite the vapors above the fluid of the test bath Damage may occur to the unit and the warranty would be void 4 Instrument Descriptions Figure 1 Instrument De...

Страница 10: ...ent to fully equilibrate before proceeding with any testing Temperature Calibration This routine allows the digital temperature controller to be calibrated to a certified thermometer In the calibration mode the display will always give 2 decimal places a Use a certified calibrated measuring device to acquire the temperature Calculate the difference between the measuring device and the Process valu...

Страница 11: ...ed with Over power Protection circuitry which prevents the unit from unsafe electrical conditions If power to the unit is lost then turn off the main power and turn it back on again The main power switch also functions as a circuit breaker 7 Maintenance WARNING Disconnect power to the unit before servicing to avoid exposure to high voltages and or temperatures which may result in personal injury o...

Страница 12: ...has been properly installed operated and maintained Koehler Instrument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replacement of the product In no event shall the cost o...

Страница 13: ...K34710 K34711 K34712 Brookfield Viscosity Liquid Bath Operation and Instruction Manual K34715 K34716 Manual 9 Notes ...

Страница 14: ...K34710 K34711 K34712 Brookfield Viscosity Liquid Bath Operation and Instruction Manual K34715 K34716 Manual 10 Notes ...

Отзывы: