background image

Low Temperature Torque Apparatus  

Operation and Instruction Manual 

 

K18854,55,64,65-Manual

    

 

-

17-

 

7  Maintenance

 

 

WARNING.

 

Disconnect  power  to  the  unit 

before  servicing  to  avoid  exposure  to  high 
voltages  and/or  temperatures  which  may  result  in 
personal injury or death. If you have any questions 
about maintaining your equipment, then please do 
not  hesitate  to  contact  the  Koehler  technical 
service department. 
 

7.1  Routine Maintenance

 

 

 For 

maximum 

refrigeration 

efficiency, 

periodically remove the front louvers and clean the 
condenser  with  either  compressed  air  or  a 
vacuum.  When  the  air  chamber  is  at  room 
temperature,  wipe  away  any  moisture  that  has 
accumulated.  This  will  minimize  frosting  of  the 
cooling coil. Change the oil in the speed reducers 
once  per  year  using  the  oil  specified  in  the 
manual.  Wash  the  test  rig  components  in  a 
suitable solvent and air dry. Do not permit wet rigs 
to  sit  out  as  the  components  will  corrode.  Wipe 
down the exterior of the cabinet using a soft cloth 
sprayed  with  household  cleaner.  Over  the  course 
of  time,  some  instrument  parts  may  need  to  be 
replaced.  When  ordering  replacement  part(s), 
please  provide  the  model  number,  serial  number, 
and  product  shipment  date  of  your  equipment  so 
that  we  can  ensure  you  will  receive  the  proper 
replacement part(s). 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 

7.2  Replacement Parts 

Part Number 

Description 

050-001-033 

Switch, Blower Interlock 

050-230-002 

Switch, Rocker Green Lens, 230V 

090-240-021 

Contactor, DPST, 30A, 240V 

091-032-001 

Relay, Solid State, 4-32 VDC, 20A 

275-103-037 

Temp Safety Limit, 230V, RTD 

278-001-002 

Fuse, 1A, Slo-Blo, 5 x 20 mm 

278-002-002 

Fuse, 2A, Slo-Blo, 5 x 20 mm 

278-020-004 

Fuse, 20A, 600VAC, Time Delay 

278-104-002 

Fuse, .25A Slo-Blo 5 x 20 mm 

301-002-007 

Timing Belt 

K18860-1-30 

Test Shaft Method D1478 

360-000-001 

Load Cell 

360-230-001 

Strain Gauge Indicator, 230VAC 

289-001-006 

Test Bearing 

289-001-009 

Ball Bearing 

Содержание K18854

Страница 1: ...mia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumen...

Страница 2: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 1...

Страница 3: ...rque Apparatus was manufactured in conformance with the applicable standards set forth in this certification Specifications ASTM D1478 ASTM D4693 ASTM D4950 FTM 791 334 This unit is tested before it l...

Страница 4: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 3...

Страница 5: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 4...

Страница 6: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 5...

Страница 7: ...information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Repo...

Страница 8: ...ely collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is gr...

Страница 9: ...1 3 3 Setup 11 4 Descriptions 12 4 1 Instrument Controls 12 4 2 Temperature Controller Operation 12 5 Operation 14 5 1 Preparation for Operation 14 5 2 Operating Instructions 14 5 3 Calibration Proced...

Страница 10: ...g needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your re...

Страница 11: ...instructions for preparing the equipment assume that the user is aware of the contents of this document which lists the warranty conditions and important precautions 3 1 Packing List Low Temperature T...

Страница 12: ...medium otherwise damage may occur to the unit and the warranty will be void It is strongly suggested that a voltage stabilizer be used to prevent voltage variability The unstable voltage may cause dam...

Страница 13: ...TM D4693 Test Rig Right side meter Over torque cut off feature The torque meter is equipped with an automatic cut off that interrupts power to the drive motor when measured torque exceeds a programmed...

Страница 14: ...g device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Press the DOWN key CAL will display on the lower gree...

Страница 15: ...followed by the outer compression plate with the shoulder facing the hub Slip on the washer and thread on the nut but do not tighten 6 Carefully mount the test assembly in a vertical position in a vi...

Страница 16: ...st Chamber After testing has been completed and the refrigeration system shuts down allow the test chamber to return to room temperature before wiping it dry with a clean cloth This will minimize fros...

Страница 17: ...rod and slip on the weight marked 1 The strain gauge should display 11 lbs Each additional weight should register an additional 11 lbs Adjust the span screw to make any correction NOTE Handle weights...

Страница 18: ...a suitable solvent and air dry Do not permit wet rigs to sit out as the components will corrode Wipe down the exterior of the cabinet using a soft cloth sprayed with household cleaner Over the course...

Страница 19: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 18 8 Wiring Diagrams 8 1 K18864 Unit Wiring 220 240V 60Hz...

Страница 20: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 19 8 2 K18865 Unit Wiring 220 240V 60Hz...

Страница 21: ...turn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Страница 22: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 21 Notes...

Отзывы: