background image

Low Temperature Torque Apparatus  

Operation and Instruction Manual 

 

K18854,55,64,65-Manual

    

 

-

11-

 

Koehler model K18500 High Temperature 
Wheel Bearing Grease Tester is suitable for this 
purpose. Alternatively, a drill press may be used 
if the proper speed, load and duration can be 
maintained 

 

Arbor Press 

 

3.2  Unpacking 

 

This  unit  comes  packed  in  a  crate.  Remove  the 
bath  from  the  crate  and  place  on  the  floor  with 
approximately  12”  of  space  between  the  back  of 
the  bath  and  the  wall  for  airflow. 

Blocking  the 

airflow will cause a malfunction to occur. 
 

WARNING:

 Be sure two or more individuals 

are available for extracting and moving the 
instrument from box to the floor. Individuals must 
lift in accordance to proper technique. See Figure 
below.  
 

 

 
 
 
Unwrap the unit and wipe the shell with a soft cloth 
to remove smudges caused by wrapping material. 
 
Very high ambient temperatures would also affect 
the efficiency of the unit. It is advisable to maintain 
the  environment  of  the  machine  below  22.78°C 
(73°F). 
 
Connect  the  line  cord  to  a  properly  fused  and 
grounded  receptacle  with  the  correct  voltage  as 
indicated in section 1.3. 

 

WARNING:

 For safety, disconnect the 

power when performing any maintenance and/or 
cleaning.

 

Do 

NOT

 turn the power on unless the 

bath is filled with the proper medium; otherwise, 
damage may occur to the unit and the warranty 
will be void. 
 

It is strongly suggested that a voltage 
stabilizer be used to prevent voltage 

variability. The unstable voltage may cause 
damage to the unit. 

 
Carefully unpack and place the instrument and 
accessories in a secure location. Ensure that all 
parts listed on the packing list are present. Inspect 
the unit and all accessories for damage. If any 
damage is found, keep all packing materials and 
immediately report the damage to the carrier. We 
will assist you with your claim, if requested. When 
submitting a claim for shipping damage, request 
that the carrier inspect the shipping container and 
equipment. Do not return goods to Koehler without 
written authorization.  

 

3.3  Setup 

 

Equipment  Placement.

  Choose  a  location  that 

accommodates 

the 

following 

dimensional 

requirements:  
 
Length (side-to-side): 5 ft. (1.53m) 
Depth  (front-to-back):  3  ft.  (1.0m)  plus  6  in.  (15 
cm) for air circulation 
 
After  uncrating,  roll  the  unit  into  position,  allowing 
6  inches  (15.25  cm)  of  clearance  in  back  of  the 
unit for air circulation, then chock in place.  
 

Power. 

Connect  the  line  cords  to  properly  fused 

and grounded receptacles with the correct voltage 
as  indicated  in  section  1.3  or  on  the  back  of  the 
unit.  
 

WARNING:

   

For  safety,  disconnect  the 

power  when  performing  any  maintenance  and/or 
cleaning.

 

Do 

NOT

  turn  the  power  on  unless  the 

bath  is  filled  with  the  proper  medium;  otherwise, 
damage  may  occur  to  the  unit  and  the  warranty 
will be void. 

 
 
 
 
 
 
 
 
 
 
 

 

Содержание K18854

Страница 1: ...mia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumen...

Страница 2: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 1...

Страница 3: ...rque Apparatus was manufactured in conformance with the applicable standards set forth in this certification Specifications ASTM D1478 ASTM D4693 ASTM D4950 FTM 791 334 This unit is tested before it l...

Страница 4: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 3...

Страница 5: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 4...

Страница 6: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 5...

Страница 7: ...information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Repo...

Страница 8: ...ely collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is gr...

Страница 9: ...1 3 3 Setup 11 4 Descriptions 12 4 1 Instrument Controls 12 4 2 Temperature Controller Operation 12 5 Operation 14 5 1 Preparation for Operation 14 5 2 Operating Instructions 14 5 3 Calibration Proced...

Страница 10: ...g needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your re...

Страница 11: ...instructions for preparing the equipment assume that the user is aware of the contents of this document which lists the warranty conditions and important precautions 3 1 Packing List Low Temperature T...

Страница 12: ...medium otherwise damage may occur to the unit and the warranty will be void It is strongly suggested that a voltage stabilizer be used to prevent voltage variability The unstable voltage may cause dam...

Страница 13: ...TM D4693 Test Rig Right side meter Over torque cut off feature The torque meter is equipped with an automatic cut off that interrupts power to the drive motor when measured torque exceeds a programmed...

Страница 14: ...g device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Press the DOWN key CAL will display on the lower gree...

Страница 15: ...followed by the outer compression plate with the shoulder facing the hub Slip on the washer and thread on the nut but do not tighten 6 Carefully mount the test assembly in a vertical position in a vi...

Страница 16: ...st Chamber After testing has been completed and the refrigeration system shuts down allow the test chamber to return to room temperature before wiping it dry with a clean cloth This will minimize fros...

Страница 17: ...rod and slip on the weight marked 1 The strain gauge should display 11 lbs Each additional weight should register an additional 11 lbs Adjust the span screw to make any correction NOTE Handle weights...

Страница 18: ...a suitable solvent and air dry Do not permit wet rigs to sit out as the components will corrode Wipe down the exterior of the cabinet using a soft cloth sprayed with household cleaner Over the course...

Страница 19: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 18 8 Wiring Diagrams 8 1 K18864 Unit Wiring 220 240V 60Hz...

Страница 20: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 19 8 2 K18865 Unit Wiring 220 240V 60Hz...

Страница 21: ...turn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Страница 22: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 21 Notes...

Отзывы: