A-61139 March 2000
6-5
Code
Operator Message Displayed
Description/Action Required
E230
Open access door to correct error
Open the access door.
Verify all operator accessible areas are
properly closed.
Press Run to continue.
E231
Press JOG, and try again
Press Jog and then press the Run key. Try
again.
E232
Close all access doors and RUN
Close all access doors.
Press Run to continue.
E234
Press JOG, and try again
Press the Jog key and then press Run. Try
again.
E235
E236
Turn main power OFF and ON
•
Turn the side panel switch off.
•
Turn the main power switch off.
•
Wait 5 to 10 seconds.
•
Turn the main power switch on.
•
Turn the side panel switch on.
If the error repeats, call service.
E238
User function string invalid
The P-key, End-of-Batch, etc., definition
made during installation is invalid. Correct the
definition using the Mode Setup Software.
E240
E241
E242
E243
E244
E245
Enter correct level of IA — refeed
Set or change the document image level or
use the Next key to enter a new image
address.
Press Run to continue.
E246
Conflicting input — what level?
Set or change the document image level to
confirm the proper document image level.
Press Run to continue.
E267
Clear error, re-enter IA
Press Clear. Use the Next key to enter an
image address.
Press Run to continue.
E280
Feeder off due to host request
System is busy. The feeder will be
reactivated when the system is no longer
busy.
Wait until the message clears and continue.