9. Electrical Connection
The dry sleeve is mounted, adjusted and tested with the pick up inside as a whole
unit, by KOBOLD.
The pick-up insert, including milled nut and spring, are screwed into the dry
sleeve by the customer (see included data sheet).
Caution: The dry sleeve must be free of foreign substance. The
spring must be able to press the pick-up insert flush with the
frontside of the dry sleeve.
The free cable ends will be connected to a display unit according connection dia-
gram and technical data on the included data sheet.
Caution: The complete sensor unit (sensor, measuring transducer
and cable) should not be located in an area of high intensity
electromagnetic field, as may come from high voltage
transmissions, electric motors, frequency converters etc,: this
could cause measuring errors or even destruction of the sensor
system.
Even though the pipe-system is filled the pick-up insert of
the pick-up can be changed without regulating a new
distance to the pole wheel.
Exchange pick-up insert E by opening milled nut R.
Caution: Dry sleeve G must not turn!
OM-…/44
OM-…/45
Further details about the pulse generators can be found in the additional
instructions: "Operating instructions for pulse generators for the OM series...
Type: /43../44../45"
OM-../43
OM-../45
OM-../44