Kobold KDF-9 Скачать руководство пользователя страница 5

Operating Instructions KDF-9/KDG-9

 

 

 

Page 5 

 

 

                  KDF-9/KDG-9 K04/0922 

  
1  Applications 

1.1   KDF-9 / KDG-9 

The flow meter can be used for measuring the flow of liquid and gaseous products in pipes. 
It shows the current flow in volume or mass per unit time. 
Applications:  
Measuring the flow of liquid and gaseous products in pipes as well as dosing, superimposing, monitoring, 
regulating and controlling them. 

1.2  KDF-9 / KDG-9 differential pressure regulator 

The flow meter can be used for stabilizing a constant flow of liquid and gaseous products in pipes. The 
flow is kept constant regardless of pressure changes during product inflow for type KDF-9 / KDG-9-...E/F 
or of a pressure change during product outflow for type KDF-9 / KDG-9-...A/B. 
Applications: Constant dosing, level measurement in open and closed vessels, N

2

 superposition of 

combustible media 

Caution:

 The devices should be used with the greatest possible caution to measure potentially hazard-

ous liquids and (especially) gases.  Precautionary measures must be taken to protect personnel and 
equipment from any potential danger or damage due to glass-tube breakage.  The plant operator is fully 
responsible for using the devices.  Where possible, we recommend the use of metal devices such as type 
KDS. 

2  Operational mode and system design KDF-9/KDG-9 

Float principle: 
The product flows through the meter vertically from the bottom to the top. The height of the float in the 
measuring tube is a measure of the flow quantity. The float is in equilibrium between the buoyant force of 
the flowing medium and the counteracting force of gravity on the float. The measured value is displayed 
on the measuring-tube scale with the upper edge or the indicator edge of the float (ball). 

2.1  KDF-9 / KDG-9 system design 

The meter consists of a conical measuring tube made of glass containing a float that can move vertically. 
The height of the float in the measuring tube reproduces the calibrated flow quantities on the measuring 
tube. 

2.2  Measuring principle for KDF-9 / KDG-9  

Differential pressure regulator 
The diaphragm of the controller is in a state of equilibrium when the pressure conditions are the same on 
both sides.  The pressure on the input side is determined by the pressure of the product; the pressure on 
the output side is determined by the pressure drop of the setting valve of the flow meter. 
If either the inlet or outlet pressure changes, the change in pressure is compensated by the built-in dia-
phragm valve - thus maintaining a constant set flow rate. 

Important:

 The controller can only regulate inlet or outlet pressure fluctuations. Steady pressure condi-

tions must prevail on the other side. 

2.3  KDF-9 / KDG-9 system design 

The unit consists of a KDF-9/KDG-9 variable-area flow meter, equipped with a diaphragm differential 
pressure flow controller. The variable-area flow meter consists of a device fitting with an integrated meas-
uring tube made of glass that contains a vertically movable float and the valve for setting the flow rate. 
The differential pressure flow controller is made of stainless steel and consists of a diaphragm made of 
Viton or PTFE and a compensating valve made of stainless steel. For gaseous products, two versions are 
available:

  

KDF-9/KDG-9-...A/B

 for a constant inlet pressure and a variable outlet pressure 

KDF-9/KDG-9-...E/F

 for a constant outlet pressure and a variable inlet pressure 

For liquids, both versions can be used; however, the 

KDF-9/KDG-9-…A/B

 version should be preferred. 

Содержание KDF-9

Страница 1: ...Operating Instructions for Flow Meter Model KDF 9 KDG 9...

Страница 2: ...DITIONS OF USE 7 6 1 Mounting start up 8 6 2 Ambient conditions 8 6 2 1 Ambient temperature ranges 8 6 2 2 Storage temperature 8 6 2 3 Climatic category 8 6 2 4 Degree of protection 8 6 2 5 Shock resi...

Страница 3: ...ating personnel 12 15 PACKAGING STORAGE AND SHIPMENT 12 16 MAINTENANCE 12 ATTENTION 13 REPLACING THE MEASURING CONE 13 REMOVAL 13 INSTALLATION 13 17 RETURNING DEVICES FOR REPAIR AND SERVICE 15 18 DECO...

Страница 4: ...avail able online may not always correspond to the product version you have purchased If you need an instruction manual that corresponds to the purchased product version you can request it from us fre...

Страница 5: ...the measuring tube scale with the upper edge or the indicator edge of the float ball 2 1 KDF 9 KDG 9 system design The meter consists of a conical measuring tube made of glass containing a float that...

Страница 6: ...values with fully opened valve Restricted contact adjustment range contact only adjustable as minimum contact Measuring ranges water 20 C Measuring ranges air 1 013 bar abs 20 C Float st 1 4401 316L...

Страница 7: ...hin 10 100 of the range 5 3 Repeatability Liquid Gas 1 0 FS E F A B option differential pressure regulator 1 5 2 5 FS within 10 100 of the range 5 4 Influence of ambient temperature none 5 5 Influence...

Страница 8: ...Void float impact e g caused by solenoid valves as this is likely to damage the measuring section or float 6 2 Ambient conditions 6 2 1 Ambient temperature ranges 20 C to 100 C with limit switch 20 7...

Страница 9: ...re regulators are not pressure reduction valves The measuring device consists of variable area flowmeter with control valve and mounted differential pressure regulator The flow amount will be adjusted...

Страница 10: ...2 Dimension drawing KDF 9 KDG 9 A B with conn NPT F as regulator c w constant inlet pressure 7 3 Weights KDF 9 KDG 9 0 4 kg KDF 9 KDG 9 E F A B 0 8 kg 7 4 Materials Fitting connections setting valve...

Страница 11: ...respect to the Pressure Equipment Directive 2014 68 EU the devices fall within the scope of appli cation of Article 4 Section 3 and need no CE mark in accordance with this directive Kobold Messring c...

Страница 12: ...ts 14 2 Installation start up and operating personnel Only trained specialists authorized by the system operator may carry out the installation electrical instal lations start up maintenance and opera...

Страница 13: ...brush and appropriate media When assembling and reinstalling the system always new gaskets have to be used Replacing the measuring cone Removal Close valve in front and behind the unit Close needle va...

Страница 14: ...ng the life cycle This requires tightening of the thrust piece 1 Loosen thread pin 2 M4x8 with hexagon 2mm and remove valve knob 3 Tighten thrust piece 1 SW14 with a torque of 3 8 Nm 4 0 Nm Caution Va...

Страница 15: ...dous materials Consequently all devices sent to us for repair must be free of any hazardous materials This also applies to possible hollow spaces and fissures in the devices If repair is necessary con...

Страница 16: ...asured medium Since this measured medium is dangerous in water poisonous corrosive flammable we have checked that all hollow spaces of the device are free of these materials neutralized and flushed al...

Страница 17: ...1 Special range on re quest on re quest KDF 92YY Gases Order Details Example KDG 9288 NV 0 00 0 Measuring range air Nl h Valve seat mm Pressure drop mbar Order no stainless steel Connection Gasket op...

Страница 18: ...gnetic compatibility EMC Part 6 3 Generic standards Emis sion standard for residential commercial and light industrial environments EN 55011 2011 Industrial scientific and medical equipment Radio freq...

Отзывы: