HND-R106
HND-R106 K09/0422
page 19
Menü
Parameter Value
Description
Calibration: Select buffer series
GPH
Technical Buffer series: GPH-Capsules
(pH 7, pH 4, pH 10)
PHL
Technical liquid buffer series: PHL ( pH 7,
pH 4, pH 10)
dIn
DIN 19266 buffer series pH 1.68(A), pH 4.01
(C), pH 6.87 (D), pH 9.18 (F), pH 12.45 (G)
Edit
Arbitrary buffer, manual input
Calibration: Calibration reminder period (factory setting:
off)
1...365
Calibration reminder period (in days)
oFF
No calibration reminder
Unit t: Select temperature unit
°C
All temperatures in degree Celsius
°F
All temperatures in degree Fahrenheit
Auto Hold: Auto measuring value identification
on
Auto measuring value identification Auto
Hold
oFF
Standard hold function on key press
Auto Power-Off: Select power-off delay
1…120
Power-off delay in minutes. Device will be
automatically switched off as soon as this
time has elapsed if no key is pressed/no
interface communication takes place.
oFF
Automatic power-off function deactivated
(continuous operation)
Universal Output
oFF
Interface off
minimal power consumption
SEr:
Serial interface activated
dAC:
Analog output activated
01,11..91
Base address for serial interface
communication