6. Mechanical Connection
6.1. Operational conditions check-up:
•
Flow volume
•
max. operational pressure
•
max. operational temperature
Attention! Exceeding prescribed ranges may cause damage to ball-
bearings and considerable measurement errors may result.
6.2. Mounting
•
Installation can take place in horizontal and ascending pipes, flow direction
from bottom to top. Flow in the direction of the arrow.
•
Pressure- and tensile loading is to be avoided on the connection threads.
Inlet and outlet piping should be secured at least 50 mm away mechanically
from the connections.
•
Check the sealing of the connections
•
The use of flat sealings is to be preferred. The sealing surfaces are shown in
section 11 - “Dimensions“