63
Configuration
Menu item
Action
Remark
Current range
Select 4-20 mA or
0-20 mA range using
keys.
Press enter to confirm.
Process variable
Select using
keys:
Cond: Conductivity
TMP: Temperature
Press enter to confirm.
Then select characteristic
(LIN/biLIN/LOG).
Selectable decades with
logarithmic setting (LOG):
Current start
Modify digit using
keys,
select next digit using
keys.
Press enter to confirm.
Entered value applies to
selected process variable/
range
If the adjusted range is
exceeded, the device
automatically switches
to the next higher range
(Autorange)
Current end
Enter value using
keys.
Press enter to confirm.
Entered value applies to
selected process variable/
range
If the adjusted range is
exceeded, the device
automatically switches
to the next higher range
(Autorange)
Assignment of measured values: Current start and current end
Output current
[mS/cm]
20
4
200
0
20
4
200
100
[mA]
Example 1: Range 0...200 mS/cm
Example 2: Range 100...200 mS/cm
Advantage: Higher resolution in range
of interest
[mA]
Output current
100
5
[mS/cm]