45
Configuration
Menu item
Action
Choices
Temp detection
during calibration
Select mode using
keys:
AUTO: Measured by
sensor
MAN: Direct input of
temperature, no measure-
ment (see next step)
EXT: Temperature speci-
fied via current input
(only if TAN E enabled)
Press enter to confirm.
(Manual temperature)
Modify digit using
keys,
select next digit using
keys.
Press enter to confirm.
Calibration mode
Select CALMODE using
keys:
AUTO: Calibration with
Calimatic buffer set rec-
ognition
MAN: Manual input of
buffer solutions
DAT: Input of adjustment
data of premeasured sen-
sors
Press enter to confirm.
(AUTO: Buffer set)
Select buffer set using
keys (see buf-
fer tables for nominal
values).
Press enter to confirm.
-00-...-10-,
(-U1-, see Appendix)
Pressing the info key dis-
plays the manufacturer
and nominal values in the
lower line.
5