6 Calibration/Adjustment
Menu Selection
Back
Calibration
Lingua
Note:
During calibration, the function check (HOLD) operating state is active for each of the calibrated
measuring channels. The outputs respond as configured.
During calibration, Stratos Multi remains in calibration mode until stopped by personnel. When cali-
bration mode is exited, a confirmation prompt is displayed to ensure that the system is ready for oper-
ation again.
Adjustment
Adjustment involves the transfer of the calibration values obtained during calibration to Stratos Multi
or the digital sensor. These calibration values are entered in the adjustment record of the Stratos Multi
for the sensor, and directly in the sensor when using digital sensors:
Menu Selection
Diagnostics
I/II [Sensor]
Cal/Adj Record [Process Variable]
NOTICE!
Without adjustment, a measuring device will supply an imprecise or incorrect output value.
For it to measure properly, Stratos Multi must be adjusted. When using analog sensors, adjustment is
required following sensor replacement.
Assigning passcodes helps to ensure that only specialist personnel with access rights are allowed to
carry out calibrations and adjustments.
The passcodes can be changed or disabled:
Parameter Setting
System Control
Passcode Entry
6.1 Memosens Calibration/Adjustment
Menu Selection
Calibration
[I] [II] Memosens …
Note:
The calibration data are saved in the Memosens sensor. This means that Memosens sensors can
be cleaned, reconditioned, calibrated, and adjusted away from the measurement location, e.g. in a lab-
oratory. Sensors in the system are replaced on-site by adjusted sensors.