CIP/SIP Counters
Note:
This function is not available for all sensor types.
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the ap-
plication, either one chemical (alkaline solution, water) or several chemicals (alkaline solution, water,
acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Note:
If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note:
A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note:
With Memosens sensors, an entry is also made in the sensor.
Settings can be changed in the
Sensor Monitoring Details
submenu:
Parameter Setting
[I] [II] [Sensor]
Sensor Data
Sensor Monitoring Details
5.9.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the Calibra-
tion menu prior to the calibration.
Calibration mode
: Presetting of calibration mode, e.g., automatic, manual, product calibration, data
entry, temperature
Further options are available depending on the calibration mode.
Automatic:
• Select calibration solution (NaCl/KCl)
Product calibration:
• Product calibration with/without temperature compensation
Cal Presettings (Admin.)
Product Calibration
Back
Calibration Mode
Conductivity
Product
Wi
II
Conductivity
Automatic
Manual
Product
Data Entry
Temperature
Calibration presettings can be changed in the
Cal Presettings
submenu:
Parameter Setting
[I] [II] … Cond
Cal Presettings