Intended Use
Stratos Eco 2405 Oxy is used for dissolved oxygen and temperature
measurement in biotechnology, pharmaceutical industry, as well as in
the fi eld of environment, food processing, and sewage treatment.
The sturdy molded enclosure can be fi xed into a control panel or
mounted on a wall or at a post.
The protective hood provides additional protection against direct
weather exposure and mechanical damage.
The device has been designed for application with amperometric
sensors, e.g. Knick SE 703 / SE 706. It provides two current outputs
(for transmission of measured value and temperature, for example),
two contacts, and a universal power supply 24 ... 230 V AC/DC,
AC: 45 ... 65 Hz.
Terminals:
Screw terminal, suitable for single wires / fl exible leads up to 2.5 mm
2
(AWG 14).
Recommended torque for the terminal screws: 0.5...0.6 Nm.
Caution!
Additional safety precautions have to be taken for operation in
hazardous locations according to CSA (CLI, DIV2, GPA,B,C,D T4,
Ex nA IIC T4) (see special document)!
WARNING!
These short instructions are intended for commissioning
the Stratos Eco 2405 Oxy transmitters. They are intended for
experienced users and do not replace the instruction manual
(on included CD-ROM or at www.knick.de).
The shipment includes a special document with safety
information.