SE715

Manual

SE715 Series Oxygen Sensors Instructions for Use

1

Connector (Memosens®)

2

A/F 19, serial number

3

PG 13.5 thread

4

Anode 

5

Cathode

6

Membrane cap

7

ZU0939 adapter

!

WARNING – Failure to observe this warn-

ing may result in serious injury

The safety alert symbol on the nameplate 

means:

Read these instructions for use, observe the spec-

ifications, and follow the safety instructions. 

The SE715 amperometric oxygen sensor is equipped 

with Memosens technology. This provides applica-

tion benefits, e.g., safe function thanks to perfect 

galvanic isolation, digital data transmission, and 

 integrated sensor diagnostics.

The plug connector is resistant to dirt and moisture. 

Calibration data, the operating time, sensor name, 

and serial number are stored in the sensor head.  

The serial number can also be found on the quality 

certificate and the packaging label.

1  Safety Instructions

The sensor contains a liquid electrolyte. Protective 

goggles and gloves should be worn when carrying 

out maintenance work, e.g., membrane or electro-

lyte replacement.

The sensor is designed for a maximum operating 

pressure of 3 bar and complies with the pressure 

equipment directive 97/23/EC, Article 3(3). 

During installation, take care not to damage the 

 plastic body or the thread (PG 13.5).

2  Intended Use

The sensor is designed for simultaneous measure-

ment of dissolved oxygen and temperature in indus-

trial processes, particularly in the fields of drinking 

water, wastewater, water purification, activation, 

 aeration control, fish farming, fishkeeping.

3  Installation and Commissioning

 •

On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick service 

team.

 •

Remove the watering cap and briefly rinse the 

 sensor with clean water.

 •

Insert the sensor into the fitting as described in the 

fitting’s user manual or mount the adapter ZU0939 

to use it with the sensor quiver of a Portavo porta-

ble meter.  

Take care not to damage the sensor 

membrane.

 •

The sensor can be operated vertically or inclined up 

to 15° off horizontal.

 •

Connect the sensor to the cable.

Note: 

Prior to commissioning, the sensor must be 

polarized on a transmitter. Polarization should last at 

least 30 minutes, after which time the sensor is ready 

for operation.

4  Operation

4.1 Calibrating the Sensor

Calibration in air is recommended for the SE715 

 sensor. Rinse the sensor with clean water and gently 

dab the membrane dry. Refer to the user manual of 

the oxygen transmitter for further details. The cal-

culated calibration data (slope, temperature, time of 

calibration) is stored directly in the sensor. Memosens 

sensors can also be pre-calibrated in the laboratory. In 

this case, on-site calibration is not required. 

4.2 Temperature Detector 

The integrated temperature detector is primarily in-

tended for automatic compensation of the measur-

ing signal, not for high-precision and safe tempera-

ture indication or control of the process temperature.
 

5  Maintenance and Cleaning

Observe the safety instructions listed above when 

performing maintenance work on the sensor.

Clean the sensor with clean water after each produc-

tion cycle and regularly in the case of severe contam-

ination.

5.1  Replacing the Membrane

Hold the sensor upright and carefully unscrew the 

membrane cap. Fill electrolyte into a new membrane 

cap (see package contents). Carefully screw the mem-

brane cap back on. Rinse the sensor with clean water.

5.2 Cleaning the Cathode

Note:

 The sanding procedure specified below should 

only be carried out if the cathode is severely contam-

inated.
Visible deposits on the cathode can be removed with 

fine sandpaper (package contents). To do so, lay the 

sandpaper on a horizontal surface and place a drop of 

water on the paper. Remove the membrane cap from 

the sensor and move the cathode very carefully in a 

circular motion over the sandpaper. Hold the sensor 

upright and exert no pressure. Rinse the cathode with 

clean water, fill a new membrane cap with electro-

lyte, and carefully screw the membrane cap onto the 

 sensor. Rinse the sensor with clean water.

5.3 Cleaning the Anode

The anode (silver/silver chloride) has a brown to violet 

color and does not require cleaning.

6  Specifications

Model code

The markings on each sensor or on the packaging 

label include the following information:

SE715/1 - MS

 

Model designation

 

  

Connector

 

   

MS: Memosens®

  

Length

 

   

1: approx. 120 mm

Further Data

Measuring range: 

0 ... 20 mg/l

Response time t

90

:

<30 s

Resolution:

20 µg/l

Signal current in air:

100 ... 130 nA

Perm. temperature range: -5 ... 45 °C
Pressure, relative:

Max. 3 bar

Membrane material:

Silicone

Temperature detector:

NTC 22 kΩ

Connector:

Memosens

Body material:

Polysulfone

Immersion depth:

Min. 30 mm

Process connection:

PG 13.5

Tightening torque:

1 ... 3 Nm

7  Package Contents

 •

Sensor

 •

Quality Certificate

 •

User Manual

 •

Maintenance kit ZU0879, consisting of:

 

2 membrane caps

 

6 ampoules with electrolyte

 Sandpaper

 

Absorbent paper

 •

ZU0939 adapter for Portavo quiver

8  Disposal

Observe the applicable local or national regulations 

for disposal.

A

ppr

ox. 120 mm

1

6

3

4

Ø 12 mm