Manual

4  Operation

4.1 Calibrating the Sensor

2-point calibration is recommended for the SE 558 

sensor. First remove the watering cap. Then dip the 

sensor successively into two different buffer solutions 

with given pH values (eg pH 7.00 and pH 4.00) and 

calibrate the pH transmitter to these buffer values. 

Please refer to the user manual of the pH transmitter 

for further details.

4.2 Temperature Detector 

The integrated temperature detector is intended for 

automatic compensation of the pH signal and not for 

any precise and safe temperature indication or con-

trol of the process temperature (see “Further Data” 

for type). 

5  Maintenance and Cleaning

Carefully rinse the sensor tip and junction with pure 

water after each operating cycle. Under no circum-

stances must measuring solution be allowed to dry 

on these parts! 

When the sensor is not in operation, store it with sen-

sor tip and junction well submerged in electrolyte 

(3 mol/l KCl). If a sensor is stored dry for a few days 

by mistake, let it soak in electrolyte for several hours 

before use.

Contaminants can be removed as follows:
Protein:

Solution of pepsin / HCl

Lime:

Dilute acids

Silver sulfide:

Solution of thiourea / HCl

Grease:

Tenside solution / lye

6  Specifications

Model Code
The markings on each sensor or on the packaging label include 
the following information:
SE 558X/ * - * MSN  Model designation 
  

Sensor connector 

   

VP: VP (VarioPin)

   

MS: MS (Memosens®)

  

Length 

   

1: 120 mm

   

2: 225 mm

Further Data
pH range

0 ... 14

Operating temperature -5 ... 100 °C
Pressure, relative

-1 ... 3 bar

Junction

Ceramic (3x)

Electrolyte

Gel with KCl reservoir

Reference system

Ag/AgCl

Sensor material

pH glass, universal use

Body material

Glass

Mounting

Pg 13.5

Temp detector

VP connector: Pt 1000

MS connector: NTC 30 kΩ

Explosion protection
SE 558... -NMSN

ATEX II 1G Ex ia IIC T3/T4/T6

SE 558... -NVPN

ATEX II 1/2 G Ex ia IIC T3/T4/T6 Ga/Gb

7  Warranty 

12-month warranty after date of shipment in the 

event of manufacturing faults.

8  Disposal

Observe the applicable local or national regulations 

for disposal.

088979

Instructions for Use for the SE 558 pH Sensors

1  Safety Information

1.1 General Information

!

 CAUTION

To ensure proper functioning, read these instruc-

tions carefully before putting the sensor into opera-

tion. Operation and servicing of the sensor should be 

carried out by specially trained personnel.

Please take extra care during installation/fitting of 

the sensor to ensure that neither the polymer body 

nor the thread (Pg 13.5) suffer any damage.

1.2 Application in Hazardous Locations

!

 WARNING

The sensor may be connected to intrinsically safe cir-

cuits. To ensure safe use, observe the electrical input 

ratings in accordance with the temperature classes 

as specified in the EC-Type Examination Certificate. 

The input ratings are safety limit values which must 

not be exceeded. 

If the stated medium temperatures are adhered to, 

no impermissible temperatures for the respective 

temperature class will occur on the equipment.  

If the sensor has been connected to non-intrin-

sically-safe circuits, it may only be reconnected to 

intrinsically safe circuits following an inspection for 

compliance with explosion protection requirements. 

The provisions of EN 60079-10 / EN 60079-14 or the 

corresponding local regulations must be met during 

installation and commissioning of the sensor in haz-

ardous locations. See the Type Examination Certifi-

cate for electrical input ratings.

The sensor may be used in constantly, long-term or 

frequently hazardous locations. Prior to installation, 

the installation site and its potential explosion haz-

ard must be assessed and the suitability of the sen-

sors determined. 

Incorrect installation or handling of the sensor may 

impair its explosion protection and result in ignition 

of an explosive atmosphere. The following points 

must therefore be strictly adhered to when using 

the sensor in hazardous locations: 

 •

In hazardous locations the sensor may only be in-

stalled by specially trained personnel authorized 

by the operating company. Installation shall be 

performed in accordance with the relevant regu-

lations and the electrical input ratings provided in 

the Type Examination Certificate. 

 •

The sensor shall be installed, commissioned and 

operated in accordance with these instructions 

for use. 

 •

Make sure that the sensor is properly mounted 

and that the O-rings are not damaged. Only this 

ensures that IP 67 protection is maintained. 

2  Intended Use

The sensors of the SE 558 series are particularly suit-

able for continuous measurement of pH values in me-

dia of low conductivity. They are low-maintenance 

pH sensors with gel reference electrolyte. The sensor 

has an integrated temperature detector for automatic 

temperature compensation of the pH signal during 

calibration and operation. It provides 3 ceramic junc-

tions and a KCl reservoir.

3  Installation and Commissioning

 •

On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick service 

team.

 •

Remove the watering cap and briefly rinse the 

sensor with clean water. After rinsing, the sensor 

should only be dabbed dry with a tissue. Do not 

rub the pH-sensitive glass, since this can lead to 

electrostatic charging and sluggish response times.

 •

Check the space behind the pH-sensitive glass for 

the presence of any air bubbles and remove them 

by gently swinging the sensor in a vertical plane.

 •

Install the sensor in the fitting as described in the 

user manual of the respective fitting.

 •

Connect sensor and cable.

pH 0 to 14, -1 to 3 bar

, -5 to 100°C

SE 558/1

1

6

3

Ø 12 mm

120 / 225 mm

4

5

8

2

9

Rating plate (here: SE 558... -NMSN)

Knick Elektronische Messgeräte GmbH & Co. KG, Beuckestraße 22, 14163 Berlin, Germany. Phone: +49 30 80191-0, Fax: +49 30 80191-200, Email: [email protected]

TA-SE558-KNX02  20150106

1

Connector (VP, Memosens®)

2

19 mm A/F, serial number

3

Pg 13.5 thread

4

PVDF compression ring 

5

EPDM-FDA O-ring (11.5 x 2.6 mm)

6

Rating plate

7

KCl reservoir

8

Junction 

9

pH glass