Knick SE 557 Скачать руководство пользователя страница 2

For Sensors with VP Connector

Certificate number: 

Marking: 

TÜV 15 ATEX 164143 X

 II 1 G Ex ia IIC T4/T6 Ga

Electrical Parameters:

Temperature class

Maximum input 

voltage U

i

Maximum input

 

current

 I

i

Maximum input

 power

 

P

i

T6

12 V

100 mA

40 mW

T4

18 V

170 mA

200 mW

 

Permissible internal capacitance

 C

i

 negligible

 

Permissible internal inductance

 L

i

 negligible

Thermal Parameters: 
For the pH Sensor Type SE 557X/*-*VPN 

[-***]

:

Temperature class Maximum input 

power P

i

Ambient temperature range 

of connecting head Ta

Permissible process 

temperature

T6

40 mW

-20 °C < Ta < +55 °C 

55 °C

T4

200 mW

-20 °C < Ta < +100 °C 

100 °C

 

For the pH Sensor Type SE 557X/*-*VPH 

[-***]

:

Temperature class Maximum input 

power P

i

Ambient temperature range 

of connecting head Ta

Permissible process 

temperature

T6

40 mW

0 °C < Ta < +55 °C 

55 °C

T4

200 mW

0 °C < Ta < +135 °C 

135 °C

Special Conditions
•  For ambient temperature range / process temperature, see parameters.

•   The temperature class depends on the ambient temperature and the input power.  

Observe the user manual.

•   Metallic process connection parts shall be connected to the local equipotential bonding.  

The intrinsically safe circuit is connected to ground.  

Along this intrinsically safe circuit, potential equalization must exist.

For Sensors with Memosens Connector

Certificate number: 

Marking: 

BVS 09 ATEX E088 X

 II 1G Ex ia IIC T3/T4/T6

Thermal Parameters:
For the pH Sensor Type SE 557X/*-*MSN

Temperature class Ambient temperature range of 

connecting head Ta

Permissible process 

 temperature

T6

-20 °C < Ta < +70 °C 

70 °C

T4 

-20 °C < Ta < +120 °C 

100 °C

T3

-20 °C < Ta < +135 °C

100 °C

For the pH Sensor Type SE 557X/*-*MSH [-***]

Temperature class Ambient temperature range of 

connecting head Ta

Permissible process 

 temperature

T6

-20 °C < Ta < +70 °C 

70 °C

T4 

-20 °C < Ta < +120 °C 

120 °C

T3

-20 °C < Ta < +135 °C

135 °C

Special Conditions
•   The cable and the sensor shall only be used within the ambient temperature range specified for 

the temperature class.

•   Do not operate the Memosens sensors under electrostatically adverse process conditions.  

Prevent strong vapor and dust streams from acting directly on the connection system.

Hazardous Areas: Electrical and Thermal Parameters

Отзывы: