4.3  Pressurizing the Fitting

During possible sensor sterilization and during nor-

mal operation, it is important that the fitting and 

electrolyte pressure is adjusted to be higher than the 

effective pressure in the reactor vessel. This reduces 

the risk of process fluid getting into the sensor. 

The optimal pressure difference value to keep the 

junctions clean and obtain acceptable electrolyte 

refill intervals has to be established empirically from 

case to case, however it should be at least 0.5 bar.

5  Maintenance and Cleaning

Regularly check the applied fitting overpressure 

during operation, and re-adjust if required. Regularly 

check the electrolyte level in the sensor. When reach-

ing the end of the bulbous shaft section (reservoir), 

refill with electrolyte up to the indication line for max. 

level. Carefully rinse the sensor tip and junction with 

pure water after each operating cycle. Under no cir-

cumstances must measuring solution be allowed to 

dry on these parts! 

When the sensor is not in operation, store it with sen-

sor tip and junction well submerged in electrolyte 

(ZU 0958). If a sensor is stored dry for a few days by 

mistake, let it soak in electrolyte for several hours be-

fore use.

6  Accessories

Connecting cable 

VP6-ST 

3 m

CA/VP6ST-003A

5 m

CA/VP6ST-005A

10 m

CA/VP6ST-010A

20 m

CA/VP6ST-020A

7  Specifications

Model Designation

The model designation, which is printed on each 

sensor and on the packaging label, includes the 

 following  information:

SE 557X /1-NMSN  Model designation (example) 

  

 

 

Sensor material 

   

 

 (see pH glass table)

  

 

 

N: Alpha glass

   

 

 H: Omega glass

  

 

 

Sensor connector 

   

 

 VP: VP (VarioPin)

   

 

 MS: MS (Memosens®)

  

 

 

Solution ground 

   

 

 N: Without

  

 

 

Length 

   

 

 1: 120 mm

   

 

 2: 250 mm

   

 

 4: 450 mm

  

 

 

Ex approval

  

 

 

X: Yes

Further Data

pH range

0 ... 14

Pressure, relative

-1 ... 6 bar

Junction

Ceramic (2x)

Electrolyte

ZU 0958

Reference system

Ag/AgCl

Mounting

PG 13.5

Temp detector

VP connector: Pt1000

MS connector: NTC 30 kΩ

pH Glass (Sensor Material)

Model Code Designation Process 

 temperature

Features

N:

Alpha glass

-20 ... 100 °C (Medium imped-

ance, universal 
glass, fluoride 
resistant) 

H:

Omega glass 0 ... 135 °C

(High impedance 
for high-tem-
perature appli-
cations, very low 
alkali error, CIP/SIP 
 capable)

8  Disposal

Observe the applicable local or national regulations 

for disposal.

Instructions for Use for the SE 557 pH Sensors 

!

WARNING –  Failure to observe this warn-

ing may result in serious injury.

The warning symbol on the rating plate 

means:

Read these instructions for use, observe the 

 Specifications, and follow the Safety Instructions. 

1  Safety Instructions

1.1 All Applications – All Sensor Variants

Hazards due to pressure, temperature, aggres-

sive media, or explosive atmosphere are pos-

sible, depending on the location of use. There-

fore, the installation, operation, and servicing of 

the sensor shall only be carried out by suitably 

trained personnel authorized by the operating 

company. 

1.2 Hazardous Areas – All Sensor Variants

Observe the corresponding local requirements 

and standards for electrical installations in 

hazardous areas. For orientation, please refer 

to IEC 60079-14, EU directives 2014/34/EU 

and 1999/92/EC (ATEX), NFPA 70 (NEC), ANSI/

ISA-RP12.06.01.

1.3 Hazardous Areas – Sensors with Memosens 

Connector

Memosens Ex sensors are marked by an 

 orange-red ring. The Memosens Ex sensors 

shall only be connected to a cable of type  

CA/MS-***X** or type CA/MS-***X**-L or to an 

intrinsically safe and certified Memosens mea-

suring cable which is identical in hardware and 

function.

1.4 Hazardous Areas – Sensors with VP 

 Connector

The sensors shall only be operated with an 

approved device in floating, intrinsically safe 

 circuits. The electrical input ratings of the 

 sensors shall not be exceeded.

2  Intended Use

Together with a suitable fitting, the sensor is used for 

continuous measurement of pH values in liquid me-

dia. The SE 557 is a high-precision sensor with liquid 

electrolyte. It has an integrated temperature detector 

for automatic temperature compensation. The sensor 

is sterilizable.

3  Installation and Commissioning

1.  On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick ser-

vice team.

2.  Remove the watering cap and briefly rinse the 

sensor with clean water. After rinsing, the sen-

sor should only be dabbed dry with a tissue. Do 

not rub the pH-sensitive glass, since this can lead 

to electrostatic charging and sluggish response 

times.

3.  Check the space behind the pH-sensitive glass for 

the presence of any air bubbles and remove them 

by gently swinging the sensor in a vertical plane.

4.  Open the electrolyte filling hole before insert-

ing the sensor into the fitting. Suitable fittings 

for operating this sensor type can be found at  

www.knick-international.com.

5.  Install the sensor in the grounded fitting as 

 described in the user manual of the respective 

fitting.

6.  Connect sensor and cable.

4  Operation

4.1 Calibrating the Sensor

2-point calibration is recommended for the SE 557 

sensor. First remove the watering cap. Then dip the 

sensor successively into two different buffer solutions 

with given pH values (e.g., pH 7.00 and pH 4.00) and 

calibrate the pH transmitter to these buffer values. 

Please refer to the user manual of the pH transmitter 

for further details.

4.2  Sterilizing

For application in sterile processes, such as fermenta-

tion, sterilize the sensor before starting the operating 

cycle. This sensor type allows the whole fermenter/

reactor unit including pH sensor to be sterilized.

1

Ø 12 mm

5

3

Knick 

Elektronische Messgeräte 

GmbH & Co. KG 

Beuckestraße 22

14163 Berlin

Germany

Phone: +49 30 80191-0

Fax: +49 30 80191-200

Email: [email protected]

TA-SE557-KNE04  201600518

1

Sensor connector: 

a: VP,  

b: Memosens®

2

Serial number

3

Filling hole

4

Maximum filling height

5

Rating plate

 e

.g

. 120 / 250 / 450 

mm

Ø 32 mm

Ø 32 mm

120 

250 

450 

mm

Position electrode

this side up

max. level

D-14163 

Berlin

BVS 09 

AT

EX E 088 

X

II 1G   Ex ia IIC 

T3/T4/T6

0044

SE 557X/*-NMS

H

pH 0 to 14

,-1 to 6 ba

r,

0 to 135 °C

2

a

b

Position el ec trode

this side up

max. level

15 

AT

EX

 164143

M

ade 
in

 Ge

rm

an

y

0044

II 1 

 Ex

 ia

 IIC 

T3

/T

4/

T6

 

14163 

Be

rlin

SE

 55

7X

/1

-N

VP

N

 p 

 -1

 to

 6 

ba

C° 

00

ot 

02

-    

p

me

T

to

 14

pH

re

l

4

Отзывы: