35
DO Function Blocks
DO 3: Unical Control
Bit
Meaning
7
6
5
4
3
2
1
0
X
Reserved
1
Probe in SERVICE position (MEASURE = 0)
1
Manual, Time control Off (Auto, Time control On = 1)
X
Reserved
X
Reserved
0
0
0
No program start
0
0
1
Program: Cleaning
0
1
0
Program: Cal 2point
0
1
1
Program: Cal 1point
1
0
0
Program: Parking
1
0
1
Program: USER 1
1
1
0
Program: USER 2
1
1
1
No program start
DO 4: LOCK Control
Bit
Meaning
7
6
5
4
3
2
1
0
0
0
Measuring mode
0
1
Enabled
1
0
Busy
1
1
Not used
X
Reserved
X
Reserved
X
Reserved
X
Reserved
X
Reserved
X
Reserved