Commissioning and Configuration
Calibration Protocols
Calibration Protocols
The protocols are transmitted in the AI TBs in binary form. The DD provides a
method for converting them to a readable format. The method can be started
with “Read protocol“ in the “Protocol“ menu of the AI TB.
Each AI TB of the FF module has a ring
buffer which can store up to three pro-
tocols. A fourth protocol would over-
write the first protocol in that TB.
With “Read protocol“, the calibration
protocol of the respective module can
be read out from the AI TB. The exam-
ple shows the pH protocol of AI TB1.
At the end of the method you can con-
firm the protocol (OK) and thus delete
it from the ring buffer. After confirma-
tion you will be informed if further
data, i.e. another protocol, are avail-
able. Without confirmation (Cancel),
you can re-read the protocol as often
as you like.