11
1) Switch off the power supply to the device.
2) Open the device (loosen the 4 screws on the front).
3) Plug the module into the slot (D-SUB connector),
see figure.
4) Tighten the module's fastening screws.
5) Connect the signal lines (see next page).
6) Check whether all connections are correctly wired.
7) Close the device by tightening the screws on the front.
8) Switch on the power supply.
CAUTION! Incorrect measurement results.
Incorrect parameter setting, calibration or adjustment may result in incorrect
measurements being recorded. Protos must therefore be commissioned by a
system specialist, all its parameters must be set, and it must be fully adjusted.
Installing the Module
CAUTION! Electrostatic discharge (ESD).
The modules' signal inputs are sensitive to electrostatic discharge.
Take measures to protect against ESD before inserting the module and wiring
the inputs.
Note: Strip the insulation from the wires using a suitable tool to prevent dam-
age.