Display text
Time [s]
01: Probe in SERVICE
02: Cleaning agent
0020 s
03: Rinse water ON
0060 s
04: Rinse water OFF
0002 s
05: Buffer I
0000 s
06: Cal Buffer 1
07: Rinse water ON
0010 s
08: Rinse water OFF
0002 s
09: Buffer II
0000 s
10: Cal Buffer 2
11: Rinse water ON
0010 s
12: Rinse water OFF
0002 s
13: Probe in MEASURE
0005 s
14: Program end
Parameter Setting: Program Flows
Cal 2point (Continuous), Media Monitoring Off
Program step
Probe: Measure
Probe: Service
Rinse water
Buffer 4.01
(Pump I)
Buffer 7.00
(Pump II)
Cleaning agent
(Pump III)
HOLD
DCS ML3
(Program running)
01
02
03
04
05
06
07
08
09
10
11
12
13
14
Stroke 1 Stroke 2
Exposure time
Calibration
with 1st buffer
Calibration
with 2nd buffer
[t]
s
Cal 2point (continuous) can also be started via a DCS input signal at input BIN2
of the Unical 9000 probe controller.
Message (display)
Calibration program