133
Menu Display
ORP adjustment
The type of reference electrode is
selected during parameter setting.
Immerse electrode in calibration
medium and wait until the ORP value
has stabilized.
Enter the nominal ORP value (bottle).
Be sure to observe the correct
reference!
(as configured)
Press “OK” to confirm.
End adjustment by pressing softkey or
enter.
ORP adjustment
Return
Reference electrode
25.6 °C
200 mV
Ag/AgCl,KCl 1m
+25.5°C
Temperature
ORP input
+200 mV
+200 mV
ORP setpoint
ORP adjustment
Abort
Reference electrode
25.6 °C
200 mV
Ag/AgCl,KCl 1m
+25.5°C
Temperature
ORP input
+200 mV
ORP setpoint
+
0
220 mV
OK
ORP adjustment
Reference electrode
25.6 °C
220 mV
Ag/AgCl,KCl 1m
+25.5°C
Temperature
ORP input
+220 mV
+200 mV
ORP setpoint
Return
End
Temperature
[°C]
Ag/AgCl/KCl
1 mol/l
[∆mV]
Ag/AgCl/KCl
3 mol/l
[∆mV]
Thalamid
[∆mV]
Mercury sulfate
[∆mV]
Temperature dependence of commonly used reference systems measured against SHE
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...