34
pH
Oxy Cond
ORP
1
2
3
4
A
Example: Logger type “In Difference”
Data Logger
Example: Logger type “Limit value”
Basic interval
Max limit
Min limit
Event interval
Pre-trigger 10 x
Interval and difference (combined)
When the delta range related to the last DIFF entry is exceeded, a new entry is cre-
ated (example: entry A) and the delta range is displaced upwards or downwards by
the delta value. As long as the measured value remains within the delta range, log-
ging is performed at the preset interval. The first DIFF entry is automatically created
when the data logger is started.
Limit value (combined)
When one of the two limit values (Min/Max) is exceeded, the data are logged as
defined by the “event interval“. Additionally, the last ten measured values before an
event are recorded (pre-trigger). As long as the measured value remains within the
limits, logging is performed at the preset “basic interval”.
Delta
Delta
Delta
Interval
Time
Measured variable
Measured variable
Time
= entry
Start logger
= entry