27
Data Logger
Starting the Data Logger using CONT
Prerequisite: Data logger is configured. Every time the meter has been switched off,
the data logger must be restarted (exception: SHOT).
Measurement
STO
Measured value is maintained
set
Logger: CONT blinks
set
Address of the last recorded value +
1 blinks
(proposed start address)
If desired: Select a start address using
.
set
The measured value is saved to the selected start address (exception: SHOT).
“… FREE MEMORY” is displayed.
“LOGGER” and “active logger type” icons are displayed.
Starting the Data Logger using START
Prerequisite: Data logger is configured. All existing entries are deleted. The start
address for saving the values is 0001. Every time the meter has been switched off,
the data logger must be restarted (exception: SHOT).
Measurement
STO
Measured value is maintained
set
Logger: CONT blinks
Logger: START blinks
set
All entries will be deleted. “5000 FREE MEMORY” is displayed.
“LOGGER” and “active logger type” icons are displayed.
Data Logger