63
Selecting a Calibration Mode for 2-/4-Electrode Sensors
Note:
• All calibration procedures must be performed by trained personnel. Incorrectly
set parameters may go unnoticed, but change the measuring properties.
Calibration can be performed by:
• determining the cell constant (cell factor for toroidal sensors) with a known
calibration solution
• entering the cell constant (cell factor for toroidal sensors)
• sampling (product calibration)
• zero calibration in air or with calibration solution (toroidal sensors)
• temperature probe adjustment
Calibration adapts the device to the individual sensor characteristics.
Access to calibration can be protected with a passcode (SERVICE menu).
First, you open the calibration menu and select the calibration mode:
Calibration with calibration solution
Calibration by entry of cell constant
Product calibration (calibration with sampling)
Temperature probe adjustment
Calibration
Selecting a Calibration Mode for Toroidal Sensors
Calibration with calibration solution
Calibration by input of cell factor
Product calibration (calibration with sampling)
Zero calibration
Temperature probe adjustment