Laboratory vacuum pumps N 8__.3 FT.40.18
Technical Data
Translation of original Operating Instructions, English, KNF 121350-121429 11/18
11
N 860.3 FT.40.18
Pneumatic performance
Max. permissible operating
pressure [bar g]
1.0
Ultimate vacuum [mbar abs.]
4
Delivery rate at atm. pressure
[l/min]*
60
Pneumatic connections
Hose connection [mm]
ID 12
Ambient and media temperature
Permissible
ambient temperature
+ 5 °C to + 40 °C
Permissible media tempera-
ture
+ 5 °C to + 40 °C
Other parameters
Weight [kg]
14.8
Dimensions : L x H x W [mm]
331 x 278.5 x 291
Maximum permissible ambient
relative humidity
80 % for temperatures up to
31 °C, decreasing linearly to
50 % at 40 °C
Max. altitude of site
[m above sea level]
2000
Electrical Data
Voltage [V]
100
115
230
Frequency [Hz]
50/60
60
50
Max. operating current [A]
4.8
2.7
1.6
Power consumption pump [W] 260
240
220
Maximum permitted mains
voltage fluctuations
+/- 10 %
+/- 10 % +/- 10 %
Fuse pump
(2x) T [A]
6.3
6.3
3.15
Protection class motor
IP54
Tab. 6
* Liters in standard state (1,013 mbar)