N 810/820/840.18 EX
Operation
Translation of original Operating Instructions, English, KNF 309515-309518 06/17
23
8. Operation
8.1. Pump
8.1.1.
Preparing for Start-up
Before switching on the pump, observe the following points:
Operational requirements
Pump
All hoses attached properly
Pump
Fan openings not blocked
Specifications of the power supply correspond with
the data on the pump’s/the electrical supply unit’s
type plate.
The pump outlet is not closed or constricted.
When using a gas ballast: No explosive or
poisonous mixtures may be produced when
ventilating the pump through the air inlet.
Tab. 15
8.1.2.
Starting
Only operate the pump under the operating parameters and
conditions described in chapter 5, Technical data.
Make sure the pumps are used properly (see chapter 2.1,
page 4).
Make sure the pumps are not used improperly (see chapter
2.2, page 4).
Observe the safety precautions (see chapter 4, page 10).
WARNING
Hazard of the pump head bursting due to excessive
pressure increase
Do not exceed max. permissible operating
pressure (see chapter 5, page 12).
Monitor pressure during operation.
If the pressure exceeds the maximum permissi-
ble operating pressure, immediately shut down
pump and eliminate fault (see chapter 10, page
31).
Only throttle or regulate the air or gas quantity in
the suction line to prevent the maximum permis-
sible operating pressure from being exceeded.
If the air or gas quantity in the pressure line is
throttled or regulated, make sure that the maxi-
mum permissible operating pressure of the
pump is not exceeded.