Installation
Installation Instructions FP 70
KNF Flodos 330352_MA_FP70_EN_00
8
3 Installation
WARNING
Danger from incorrect installation
Injuries or damage to property from leaking media
➢
After installation, check the system for leaks with a safe
medium.
➢
Observe the requirements for intended use.
➢
Take precautions in the design of the end device so that
leaking liquids cannot come into contact with live compo-
nents.
➢
Use only connections that are specified in the data sheet.
G1/8’’ thread is recommended. Do not use NPT threads or
sealing tape, as cracks can occur at the plastic connec-
tions.
The following requirements apply to all activities described in this
section:
3.1 Installation location
The installation location must:
▪
protect the pump against immersion
▪
keep aggressive and flammable liquids and vapors away
▪
allow hoses to be connected without tensile or bending stress
▪
take account of cooling the pump drive
▪
ensure that any leaks can be handled
▪
have appropriate precautions if flammable media are used
Mounting orientation
The following points are recommended for optimal filling and
venting of the pump head.
▪
horizontal mounting position
▪
Filling / emptying pump with nominal rpm (nominal voltage).
Figure 1: Horizontal orientation