
55
Table of Figures
Table of Figures
Fig. 1:
Overview valves and valve drives . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
Fig. 2:
Diagram of a multiposition valve 7P/1C . . . . . . . . . . . . . . . . . . . . . .
7
Fig. 3:
Injection valves 6P/2C and 6P/3C . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
Fig. 4:
Valve positions loading and injecting . . . . . . . . . . . . . . . . . . . . . . . .
8
Fig. 5:
Column selection – injection from the 6P/3C valve on column A . . .
Fig. 6:
Column selection – injection from the 6P/3C valve on column B . . .
Fig. 7:
Column selection with a 2-channel valve . . . . . . . . . . . . . . . . . . . .
Fig. 8:
Backflushing – injection from the 6P/3C valve on the column . . . . .
Fig. 9:
Backflushing – switchover of the 6P/3C valve . . . . . . . . . . . . . . . . .
Fig. 10:
Backflushing precolumn – injection from the 6P/3C valve on
the precolumn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 11:
Backflushing precolumn – switchover of the 6P/3C valve . . . . . . . .
Fig. 12:
Sample accumulation on precolumn – injection from the
6P/3C valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 13:
Sample accumulation – analysis of the accumulated sample . . . . . .
Fig. 14:
Precleaning of a sample by a precolumn . . . . . . . . . . . . . . . . . . . . .
Fig. 15:
Precleaning of a sample – Switching over to main column . . . . . . .
Fig. 16:
Application example right: Precleaning alternating on two
precolumns and analyze all substances from both precolumns. . . . .
Fig. 17:
Alternating sample precleaning – sample is filled in the
precolumn B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 18:
Alternating sample precleaning – precolumn B is purged and
precolumn A is filled with sample . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 19:
Device front side of the valve drive with mounted injection valve . .
Fig. 20:
Rear view of the device, with connections . . . . . . . . . . . . . . . . . . . .
Fig. 21:
Exploded view of an electrical valve . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 22:
Back side of the electrical valve – adapter plate for mounting . . . . .
Fig. 23:
Mounting of the valve on the valve drive . . . . . . . . . . . . . . . . . . . .
Fig. 24:
Mounting plate for valves and column holder . . . . . . . . . . . . . . . . .
Fig. 25:
Biconical seal and Dynaseal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 26:
Biconical seal, capillary and screw fitting as example of
the syringe connection (A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 27:
Injection of a sample with glass syringe and metal cannula . . . . . . .
Fig. 28:
Example of a multifunction module (‘Assistant’) . . . . . . . . . . . . . . .
Fig. 29:
Terminal strip RS-232, LAN, and remote control
. . . . . . . . . . . . . . . 32
Fig. 30:
Remote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Fig. 31:
Indication of the injection position on the valve's display . . . . . . . .
Fig. 32:
Maximum number of injection positions
. . . . . . . . . . . . . . . . . . . . . 37
Fig. 33:
Display, load sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 34:
Injection syringe with luer lock and flat-ground cannula 1/16” . . .
Fig. 35:
Display, inject sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 36:
Connecting the flat ribbon cable with the connector strip . . . . . . . .
Fig. 37:
Remote
and
Interface Box IF2 . . . . . . . . . . . . . . . . . . .
41