37
Operation
BlueShadow Detector 40D Instructions V7647
5.
Move the cursor to the double arrow on the lower left line to navigate
to the next line of the program or to create a new line .
6.
To create a new line, set the start time for the new program step using
the vertical arrow keys
or the confirmation key
combined
with the numeric pad . Use the confirmation key
to continue with
the setting of wavelength and events of this program line (use ho-
rizontal keys
to switch between parameters as well as between
wavelength and events screen) .
5.6.1 Creating a program
The use of the display is demonstrated using this exemplary program:
Program 01 is used to monitor absorbance at 280 nm for the first
10-minute step and at 420 nm during the next step, which has no time
limit, since it is the final step in this example . Additionally, event 1 is ena-
bled during the second step .
Process
Figure
1.
Select the menu Device Programs .
2.
Select a program number (from the currently
displayed screen) .
3.
Adapt time-, wavelength- and event-parame-
ters by navigating via horizontal keys
and
editing via vertical keys
or via usage of
confirmation key
and numeric keypad .
4.
Navigate to the double arrow on the lower left
side of the display and press the downward
key
to create a new line .
5.
Edit the time value of the new line and confirm
via the confirmation key
.
6.
Edit time, wavelength and events as done be-
fore (see also step 3) .
7.
Press the menu key
for 2 seconds to exit the
submenu . The program is automatically saved .