background image

Process

Figure

1. 

Place the valve head on the front  

of the mounting bracket so that the 

mark 

5

points upwards and the  

groove 

6

points to the left.

 

Fig. 3: 

Rear view valve

Process

Figure

2. 

Mount the feather key 

7

pointing to 

the AZURA® S system onto the groove 

6

.

3. 

Mount the adapter plate 

8

with the 

cable pointing down onto the feather 

key 

7

and tighten it with the screw-

driver size 3 mm.

4. 

Tighten the lever with the screwdriver 

size 2.5 mm and mount the lid onto 

the lever.

Fig. 4: 

Mounting a valve

Fig. 5: 

Front view

Note: 

In open position, the 6-port valve is connected: port 1 with port 6, port 2 with  

port 3 and port 4 with port 5.

5

6

7 8

Supplement AZURA® Mounting Bracket S, V6848, version 2.0 