Process
Figure
1.
Place the valve head on the front
of the mounting bracket so that the
mark
5
points upwards and the
groove
6
points to the left.
Fig. 3:
Rear view valve
Process
Figure
2.
Mount the feather key
7
pointing to
the AZURA® S system onto the groove
6
.
3.
Mount the adapter plate
8
with the
cable pointing down onto the feather
key
7
and tighten it with the screw-
driver size 3 mm.
4.
Tighten the lever with the screwdriver
size 2.5 mm and mount the lid onto
the lever.
Fig. 4:
Mounting a valve
Fig. 5:
Front view
Note:
In open position, the 6-port valve is connected: port 1 with port 6, port 2 with
port 3 and port 4 with port 5.
5
6
7 8
Supplement AZURA® Mounting Bracket S, V6848, version 2.0
Page 2