Knauer Azura GPC Скачать руководство пользователя страница 3

Operating modes

Positions for column bypass / manual injection valve

16 Port Multiposition valve

Inject (2)

Load (1)

6 Port 2 Position valve

1

1

2

2

3

3

4

4

5

5

6

6

A

B

C

3

Sample loop selection valves / fractionation valve

1 2

3

4

16

A. Loading samples

1.  Set the injection valve to Load (1) position.
2.  Address the sample loop you want to load 

(start with position 1) by switching the two sam-
ple loop selection valves.

3.  Rinse the sample loop tubing with solvent. Op-

tionally empty tubing by injecting air. 

4.  Load the individual sample through the injec-

tion port of the injection valve with 3x excess of 
the sample loop volume.

5.  At the end of sample loop loading, switch the 

two sample loop selection valves to the next 
position. The loaded loop is now closed.

6.  Rinse the tubing and valves with solvent. Optio-

nally empty tubing by injecting air.

7.  Repeat procedure for all required sample 

loops.

8.  After sample loop loading: Switch the two 

sample loop selection valves to waste (position 
16 with short tubing) and rinse with solvent to 
remove air. 

Note: 

For each loading, make sure that the 

same position is set for both sample loop se-
lection valves. If both valves are not switched 
simultaneously, the fl ow must be stopped. 
Otherwise strong pressure spikes can occur 
and may cause the pump to stop.

Note:

 To prevent air bubbles to enter the 

sample loop do not remove the syringe 
before switching both sample loop selection 
valves to next or washing position (position 
16 with short tubing).

B. Equilibrating column

Before the sample cleanup the GPC column has to 
be equilibrated with the desired eluent. Procedure:
1.  Set the injection valve to Load position. 
2.  Set column bypass valve to Inject position to 

integrate the column into the fl ow.

3.  Switch fractionation valve to designated Waste 

position.

Note: 

It is suggested for the column equilibrating 

step to use position 16 of the fractionation valve 
as waste.

C. Running the GPC

After sample loops have been fi lled, the samples 
are processed one after the other. Procedure:
1.  Set the injection valve to Inject position. 
2.  Select the individual sample loops by setting 

the two sample loop selection valves.

3.  Set the column bypass valve to Inject position 

to integrate the column into the fl ow.

4.  Switch the fractionation valve to collect frac-

tions.