Unpacking and Setup
11
DG 2.1S user manual V6880, version
3.3
Unpacking and Setup
Contacting the Technical Support
You have various options to contact the Technical Support:
Phone: +49 30 809727-0
Fax: +49 30 8015010
E-mail: [email protected]
You can make your requests in English and German.
Location Requirements
Requirements
The location for the device must meet the following requirements:
Device defect
The device overheats at exposure to sunlight and insufficient air circulation.
Device failures are very likely.
Set up the device in such a way that it is protected against exposure to direct
sunlight.
Keep at least 15 cm clear at the rear and 5–10 cm at each side for air circula-
tion.
General warning sign that moderate injuries or device
defects can occur
Device fulfills the requirements of the Conformité
Européenne, which is confirmed by the Declaration of
Conformity.
Weight
2.3 kg
Dimensions
121 × 138 × 190 mm (Width × Height × Depth)
Power supply
85–265 V
Air humidity
< 90 %, non-condensing
Temperature
4-40 °C,39.2-104 °F