48
Maintenance and care
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
9.1.2 Maintenance contract
Opening the device or removing housing parts may only be performed
by KNAUER or a company authorized by KNAUER and is covered by a
sepa rate maintenance contract.
9.1.3 Maintenance intervals
Using Mobile Control or software, you can read out the operating hours
of the detector. You find a detailed description on how to read out GLP
data in chapter „5.6 GLP“ on page 32.
Operating hours
Measures
1000
Replace halogen lamp (DAD 6.1L).
2000
Replace deuterium lamp.
6000
Replace flow cell.
Replace fiber optic cables.
9.2 Cleaning and caring for the device
9.2.1 Handling the flow cell cartridge
An important aspect in handling your flow cell cartridge, is that you
should not touch the fiber optic ends with your fingers. Your finger may
leave a thin fat layer on the fiber optic ends, which drastically impairs the
flow cell’s and detector’s performance (see section „8.2 Possible prob-
lems and solutions“ on page 43).
To diagnose this issue, we recommend generating an intensity spectrum
(via your chromatographiy software under Diagnostics). Dirty fiber optic
ends result in little or no UV light (see below).
Operating hours