DAD 6.1L/DAD 2.1L/MWD 2.1L user manual V6700, version 3.2
26
Operation
You find a detailed description on chromatography software in the corre-
sponding user manual.
Operating with Control Unit
The Control Unit is an optional accessory. To control the device using the
Control Unit, connect the cable of the Control Unit to the device. You find a
detailed description on the Control Unit in its accompanying user manual.
Operating with Mobile Control
The Mobile Control is a device control software which can be installed on
your computer or tablet. To control the device using the Mobile Control,
connect the computer or tablet with a Windows operating system to a wire-
less LAN router. The firmware version of your DAD 6.1L must be 01.21 or
higher, and 01.01 or higher for the DAD 2.1L or MWD 2.1L. Data transfer
between device and app is actualized through wireless LAN. You find a
detailed description on the Mobile Control in its accompanying user man-
ual.
Meaning of the LEDs
There are three LEDs and a switch on the front of the device.
The LEDs can have different colors depending on the operating conditions.
Standby
To start the standby, keep the standby button pressed for 5 seconds.
Note:
Malfunctioning system after repeated standby possible. After repeat-
edly using standby, restart the device using the power switch to reset the
device's data storage.
Legend
1
Left LED
2
Center LED
3
Right LED
4
Switch/Standby
button
Fig. 21
LEDs and switch/standby button on the front of the device
1 2 3
4
Color
Status
Operation
Left LED
red
Error
Check the system.
Shortly press the
switch to deactivate
the error message.
green
3D data are acquired.
Center LED
does not light
The lamp has been
switched off.
flashes green
The lamp/lamps are
initializing or the vali-
dation is progressing.
Wait until the lamp
is running or the
validation is fin-
ished.