Technical Data are subject to change without notice!
Safety notes
•
The device should only applied by instructed personnel.
•
The power head emits strong magnetic radiation at 30-60 kHz to a distance of 300 mm.
Therefore persons with cardiac
pacemakers
should
not work
with this device!
•
Magnetic data storage media should be kept in a distance of at least 3m from the power
head to avoid data loss. The same is valid for electromagnetic sensitive parts, devices and
systems.
•
Do
not place
the power head in the switched-on state
on metallic objects
, because this
results in eddy currents which could overload the device and strong heat up small objects.
Also the probe could be destroyed!
•
No metallic objects, other than the disc-type coil, should be located in the air gap of the
power head. The same applies to metallic parts within a radius of up to 50 mm in all
directions.
•
Do not use damaged or faulty cables!
•
Never touch in the area between shaft and inductive head, the rotating shaft itself or rotor
electronic contacts during operation!
•
This is a “Class A” system suitable for operation in a laboratory or industrial environment.
The system can cause electromagnetic interferences when used in residential areas or
environments. In this case the operator is responsible for establishing protective
procedures.
Содержание MTP-NT
Страница 4: ...Version 2018 14 CN 4 Technical Data are subject to change without notice MTP NT connection overview ...
Страница 6: ...Version 2018 14 CN 6 Technical Data are subject to change without notice MTP NT Bock diagram ...
Страница 21: ...Version 2018 14 CN 21 Technical Data are subject to change without notice ...
Страница 25: ...Version 2018 14 CN 25 Technical Data are subject to change without notice ...
Страница 64: ...Version 2018 14 CN 64 Technical Data are subject to change without notice Dimensions of IND PWR HEAD XXL ...
Страница 65: ...Version 2018 14 CN 65 Technical Data are subject to change without notice Dimensions of IND PWR HEAD XXL ...