background image

4

Maintenance, Service, Warranty

Maintenance:

the SMC 133 power analyzer does not require any additional maintenance during its operation

besides the regular maintenance processes for the whole cabinet. For reliable operation it is only necessary to

meet the operating conditions specified and not expose the instrument to violent handling and contact with

water or chemicals which could cause mechanical damage.

The lithium cell built in the instrument can backup a real time circuit for more than 5 years without power

supply, at average temperature 20

°

C

and load current in the instrument less than 10

µA

. If the cell is empty, it

is necessary to ship the instrument to the manufacturer for battery replacement.

The 18350 rechargeable Li-ion battery cell for the instrument operation backup is also optionally installed

inside the SMC 133 option S or L. If the battery malfunctions it is necessary to ship the instrument to the

manufacturer for battery replacement.

Service:

in the case of failure or a breakdown of the product, you should contact the supplier at their address:

K M B Systems, s. r. o.

r. dr. M. Hor´

akov´

e 559

460 05 Liberec 7

Czech Republic

Tel. 485 130 314

E-mail: [email protected]

Web:

www.kmb.cz

The product must be in proper packaging to prevent damage during transit. A description of the problem

or its symptoms must be delivered together with the product.

If a warranty repair is claimed, the warranty certificate must be sent in. In case of an out-of-warranty repair

you have to enclose an order for the repair.

Warranty certificate:

warranty period of 24 months from the date of purchase is provided for the instrument,

however, no longer than 30 months from the day of dispatch from the manufacturer. Problems in the warranty

period, provably because of faulty workmanship, design or inconvenient material, will be repaired free of charge

by the manufacturer or an authorized servicing organization.

The warranty ceases even within the warranty period if the user makes unauthorized modifications or changes

to the instrument, connects it to out-of-range quantities, if the instrument is damaged due to ineligible or

improper handling by the user, or when it is operated in contradiction with the technical specifications presented.

Type of product:

SMC 133 . . . . . . . . . . . . . . . . .

Serial number:

. . . . . . . . . . . . . . . . . . . . . . . . . . .

Date of dispatch:

. . . . . . . . . . . . . . . . . . . . . . . . . . .

Final quality

inspection:

. . . . . . . . . . . . . . . . . . . . . . . . . . .

Manufacturer’s seal:

. . . . . . . . . . . . . . . . . . . . . . . . . . .

Date of purchase:

. . . . . . . . . . . . . . . . . . . . . . . . . . .

Supplier’s seal:

. . . . . . . . . . . . . . . . . . . . . . . . . . .

15

Содержание SMC 133

Страница 1: ...cription 1 2 Operating the Meter 2 2 1 Safety requirements when using SMC 133 2 2 2 Used symbols 2 2 3 Installation of the instrument 4 2 3 1 Supply voltage 4 2 3 2 Measured voltage 5 2 3 3 Measured c...

Страница 2: ...nfrastructure and also remote load management Two different user interfaces are offered SMC 133 D with a small embedded color display on the front panel for simple local value reading and an unobtrusi...

Страница 3: ...nd this person must know in detail the operation principles of the equipment listed in this description When the device is being connected to the parts which are under dangerous voltage it is necessar...

Страница 4: ...quipment Earth ground To identify an earth ground terminal in cases where neither the symbol 5018 nor 5019 is explicitly required Class II equipment To identify equipment meeting the safety requiremen...

Страница 5: ...d or the instrument might overheat and or the internal temperature value measured may be influenced A connection wire s maximum cross section area is 2 5 mm2 in case of all screw terminals The SMC 133...

Страница 6: ...urrent transformers must be con nected to the terminal pairs I11 I12 I21 I22 I31 I32 with a cable of maximum cross section area 2 5 mm2 and a maximum length of 3 m X 100mA current input option The sup...

Страница 7: ...surements star delta and single phase feeders Option E left for Ethernet port remote communication option U right for local USB communication port all instruments provide RS485 serial line Option S 10...

Страница 8: ...ernal power supply can damage the instrument DI O as a digital input DI is connected to the selected I O terminal D1 or D2 and a common terminal C1 or C2 with external power supply in the following wa...

Страница 9: ...to download The actual download will start by pressing the Download All button When finished the archive can be viewed in the ENVIS application User can open the downloaded file directly from ENVIS D...

Страница 10: ...N A DC auxiliary voltage range power supply 3 5 W 7 VA 3W overvoltage category III 50 V II 100 V III 300 V II 600 V internal backup power supply adjustable 1 3600 s N A pollution degree 2 maximum ope...

Страница 11: ...supply backup battery Li ion 18350 charging deactivated under 0 C protection class II weight max 0 25 kg spec Other en EN 61000 4 2 6 kV 8 kV EN 61000 4 3 10 V m up to 3 GHz EN 61000 4 4 4 kV EN 6100...

Страница 12: ...certainty 0 3 of rdg or 0 3 THDU measuring range 0 20 measuring uncertainty 0 5 reference conditions measuring range measuring uncertainty twice the levels of class II acc to IEC 61000 4 7 ed 2 spec U...

Страница 13: ...measured value affected by the instrument power dissipation measuring range 40 80 C measuring uncertainty 2 C spec IT en INOM IB rated basic current 0 1 AAC 5 AAC 0 00025 0 15 AAC 0 0125 7 5 AAC intr...

Страница 14: ...UNOM x INOM reference conditions A ambient temperature tA U I for active power PF cos for reactive power 23 2 C U 80 120 UNOM I 1 120 INOM PF 1 00 PF 0 00 0 5 of rdg 0 005 PNOM PF cos uncertainty 0 0...

Страница 15: ...al Inputs type Opticaly isolated unipolar Cx Dx maximum voltage voltage for logical 1 voltage for logical 0 input current spec IO en 35 VDC 30 mADC dynamic parameters pulse output pulse duration gap d...

Страница 16: ...tems s r o Tr dr M Hora kove 559 460 05 Liberec 7 Czech Republic Tel 485 130 314 E mail kmb kmb cz Web www kmb cz The product must be in proper packaging to prevent damage during transit A description...

Отзывы: