38
Revision 5.5
5.7.2
Mounting the Focusing Handpiece
WARNING
Risk of injury to eyes by laser light!
•
Never start device as long as the protective caps cover the beam path.
•
Never activate MCO 25 plus laser without focusing handpiece. The laser light that is
emitted from the mirror joint arm has a divergence that is so low that even in a distance
of several meters the laser beam is still focused, which bears significant hazard potential.
Preparing the arm for the focusing handpiece
•
Ensure that the lever (2) is set to
handpiece
.
—
Note: the lever (2) is set to
handpiece
by factory default.
•
Unscrew the protective plastic cap that is located on the end of the mirror joint arm,
see Fig. 5-4.
•
Screw on a selected focusing handpiece, see section 12 “Accessories”, page 105, finger-
tight.
•
After connecting the focusing handpiece, insert the transparent flushing air hose into the
hose connection of the focusing handpiece, see red circle in Fig. 5-6.
Fig. 5-6:
Connection for flushing air hose,
example: focusing handpiece with a focal length of 127 mm