![KLS Martin marLED X Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990056.webp)
56
Switching on the receiver unit
•
Energize the receiver unit via the power supply unit (power supply connection (8, Fig. 7-1)).
•
Switch on the receiver unit (on/off switch (7, Fig. 7-1)).
–
The green POWER LED (4, Fig. 7-1) lights up.
Pairing of camera and receiver unit
This process is required only in order to form a new pair of camera and receiver unit.
•
Connect the monitor to the receiver unit and ensure that the monitor is switched on.
•
Switch on the receiver unit, see “Switching on the receiver unit”, page 56.
•
Switch on the light/surgiCam® camera.
•
Press and hold the
CHANNEL
(6, Fig. 7-1) button on the receiver unit for more than 10 s.
–
The
STATUS
LED (3, Fig. 7-1) goes dark and then starts flashing slowly (channel search).
–
When the
STATUS
LED is lit yellow, the connection has been established (successful pairing).
•
Repeat this process if necessary.
•
Please contact the Martin Service Center if the connection cannot be successfully established after 2–3
attempts.