![KLS Martin marLED X Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990027.webp)
27
5.1.1
Cable Specifications
All cables used for the Gebrüder Martin CAT system must meet the CAT5 specification along with the
requirements of the TIA-568 B standard.
Possible malfunction of the light system!
Using RJ45 cables that do not comply with the above-mentioned specification can lead to mLB failure.
•
Make sure that both cable ends are identically connected to the respective plug.
•
Use only 1:1 cables – do not use crossover or host cables!
Cables supplied by Gebrüder Martin comply with the TIA-568 B standard, see section 5.1.2 “Pin
Assignment acc. to TIA-568 B”, page 27.
5.1.2
Pin Assignment acc. to TIA-568 B
Assignment when replacing a connector
•
The 4 twisted-pair wires P1 to P4 of a CAT5 cable are
assigned to the plug-in contacts 1–8 as shown on the
left.