14
Manual CPM 374
If the supply voltage of the NiMH battery falls below 6.7 Volt, the device returns in the upper
line:
The battery has to be charged, although the current measurement can be completed.
Important:
Please use only the provided power supply to charge the battery. Charging also
takes place while the device is switched off. The maximum charge time for the switched off de-
vice is about 14 hours.
If the supply voltage falls below 6.4 Volt, the display shows
before the device powers off automatically to prevent total discharge of the battery. Total dis-
charge can lead to damage or breakdown of the battery!
8.
Warranty
In case of proper operation according to the manual we issue a guarantee of 24 months. Mechan-
ical damage and battery destruction by wrong handling are excepted.
Guarantee expires on opening the device!
9.
Grounding
To provide accurate evidence of the polarity and magnitude of the field measured, the device has
to be grounded properly. To this end connect the instrument at the grounding jack with ground.
10.
Warning Notices
•
The instruments are not accredited for measurements in explosion-proof areas. Mea-
surements in power plants are not permitted!
•
Alternating fields f > 1 Hz cannot be measured
•
Spark discharge on the modulator system must be avoided
•
Do not connect the switched off device to the mains adapter for longer periods (max.
14h), overloading may occur
AUTO OFF
LOW BATTERY !
LOW BATTERY !
EFM DEFECT !
REPAIR IS NEEDED